10
AirAware
005
3. Press and hold both the
and
keys on instrument faceplate
simultaneously for 5 seconds. "Bump" will appear in the display indicating
the instrument is now in the Set-up Mode. At this point the alarm relays*
are disabled, audio alarms* are disabled, and the current output* locks at
3.0 mA for all gases except oxygen, which goes to 16.0 mA until set-up is
complete. If no buttons are pressed within a 3-minute time period, the
instrument reverts to normal operation.***
006
4. Pressing the
key will advance forward through the various options,
while pressing the
key will step backward. The internal keys act as
or
buttons (Page 9) to set the actual set-up values or parameters.
*
Indication of purchase option.
**
Will not appear on menu if option not purchased.
***
A flashing decimal point on the lower right corner of the display will appear for 3 minutes after start-up,
calibration, bump, or programming. This is an indicator that the 4-20 mA signal is locked at 3 mA and that all
alarm functions are disabled. Once this flashing indicator clears all functions will return to normal.