Valor™ 2000W Series

                                                                           

EN-5

 

 

TABLE 3-2 Display Symbols

 

Item  Description 

Item

Description

1  NET symbol 

Dynamic symbol* 

2  Center of Zero symbol 

Battery charge symbol 

3  Negative symbol 

Pound, Ounce, Pound:Ounce symbol 

4  Stable weight symbol 

10 

Gram, kilogram symbol 

5  Preset Tare symbol* 

11 

Pieces symbol* 

6  Pointer symbols* 

12 

Percent symbol* 

Note:

 * Not Used 

 

                                 Figure 3-3. Valor 2000W – V22PW

 

Rear display.

 

 

3.2 Turning Scale On/Off 

To turn the scale on, press the 

On/Zero Off

 button. The scale performs a display 

test, momentarily displays the software version, and then enters the active weighing 
mode.  
To turn the scale off, press and hold the 

On/Zero Off

 button until OFF is displayed. 

 

3.3 Calibration Menu  

To enter the calibration menu, extend press 

Cal

 button for 5 seconds, [

SpaN

] will 

then be displayed. Press 

Zero

 key to perform span calibration or press 

Cal

 key to 

proceed to GEO settings. Press 

Zero

 key to enter GEO settings and then press 

Cal

 

key to increase the value. Press 

Zero

 key to confirm, [

end

] will be displayed. Press 

Zero key to exit the menu and return to weighing mode.  
  

Cal  

 

 

 

Figure 3-4. Menu structure 

 

Span

 

[

SpaN

] (yes, no) 

 

Initiates a span calibration procedure (zero and span). 

 

GEO

 

[

GEO

]

  

Geographical Adjustment Factor (GEO) is used to adjust the calibration based on 
the current location. Settings from 0 to 31 are available with 12 being the default. 
Refer to table 4-2 to determine the GEO factor that corresponds to your location. 

Span 

GEO 

End 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for Valor 2000W Series

Page 1: ...ie Valor 2000W Manual de instrucciones S rie Valor 2000W Manuel d instructions Valor 2000W Serie Bedienungsanleitung Serie Valor 2000W Manuale di istruzioni www GlobalTestSupply com Find Quality Produ...

Page 2: ...conditions specified in these instructions Service should be performed only by authorized personnel Only use weights within the scale s capacity as specified in these instructions Do not operate the s...

Page 3: ...2 Level Indicator 2 5 Power The AC Adapter is used to power the scale when battery power is not needed First connect the AC Adapter plug to the scale input jack located at the bottom of the scale acco...

Page 4: ...battery operation the battery symbol indicates the battery charging status When charging the symbol will blink slowly and when fully charged the symbol will stop blinking TABLE 2 1 Symbol Charge Level...

Page 5: ...ting Short Press Enter clear a Tare value Extended Press3 Initiates Calibration procedure Short Press in Menu Toggles through available settings Notes 1 Short Press Press less than 2 5 seconds 2 Long...

Page 6: ...s displayed 3 3 Calibration Menu To enter the calibration menu extend press Cal button for 5 seconds SpaN will then be displayed Press Zero key to perform span calibration or press Cal key to proceed...

Page 7: ...ow instructions to enter the service menu to set the displayed unit With the scale off and LFT off extend press both the On Zero and Tare keys at the same time for more than 8 seconds until rAmMP is d...

Page 8: ...oggles through active Units Short Press in Menu Toggles through available settings Short Press Initiates an application mode specific response Long Press Selects active Mode Short Press in Menu return...

Page 9: ...nstalled it should be calibrated to ensure accurate results Before performing the calibration be sure to have the appropriate calibration weights as listed in table 3 3 Press and hold Menu until mMeNU...

Page 10: ...the Low limit value Press Yes to accept or No to edit the Low limit value The stored value then displays with the first digit highlighted 000 000 kg Repeatedly press No until the desired number appear...

Page 11: ...cates factory default setting Note The Calibration Menu Mode indicator is displayed when in the Menu Mode 4 6 2 Cal Menu Enter this menu to perform calibrations Span SpaN yes no Initiates a span calib...

Page 12: ...40 05 15 15 14 14 13 13 12 12 11 11 10 40 05 41 12 16 15 15 14 14 13 13 12 12 11 11 41 12 42 19 16 16 15 15 14 14 13 13 12 12 11 42 19 43 26 17 16 16 15 15 14 14 13 13 12 12 43 26 44 32 17 17 16 16 15...

Page 13: ...rtment 5 Remove the plastic pan 6 Clean the plastic pan 6 3 1 2 4 5 Figure 5 1 Pan cleaning 5 2 2 Reinstalling the pan after cleaning Please follow the steps below to reinstall the plastic pan see als...

Page 14: ...of range Tare value exceeds maximum Err 9 Data Internal data error Contact an authorized service agent Err13 mMEmM Fail to write EEPROM Contact an authorized service agent Err 53 C SUmM Invalid checks...

Page 15: ...lay 2 x Red LED front and rear 6 digit 7 segment 20 5 mm 0 8 in characters Keyboard V22PW 2 buttons V22XW 4 buttons Application Modes V22PW Weighing V22XW Weighing Checkweighing Battery Operating Time...

Page 16: ...irective 2006 66 EC introduces new requirements from September 2008 on removability of batteries from waste equipment in EU Member States To comply with this Directive this device has been designed fo...

Page 17: ...standard s requirements On June 21 2012 OHAUS Corporation USA was re registered to the ISO 9001 2008 standard Limited Warranty OHAUS products are warranted against defects in materials and workmanship...

Reviews: