background image

Valor™ 2000W Series

                                                                         

EN-13

 

 

5.3 Troubleshooting 

The following table lists common problems and possible causes and remedies.  
If the problem persists, contact OHAUS or your authorized dealer. 

 

TABLE 5-1 

Symptom  

Possible Cause  

Remedy 

Cannot turn on  

No power to scale 
Battery discharged 

Verify connections and power 
source 

Poor accuracy 

Improper calibration  
Unstable environment  

Perform calibration  
Move scale to suitable location  

Cannot calibrate 

Unstable environment  
Incorrect calibration weight  

Move the scale to suitable location  
Use correct calibration weight 

Err 3.0  Cal

 

Incorrect calibration weight  See table 3-4 for correct weights 

Err 8.1 

 

¯LOad¯

 

Power on zero range 
exceeded 

Clear pan 

Err 8.2  _LOad_

 

Power on zero under range  Install pan 

Err 8.3 

 

¯LOad¯

 

Overload  

Load exceeds scale maximum 
capacity 

Err 8.4  _LOad_

 

Under load 

Reading below min. range - Re-
install pan. 

Err 8.5  ¯tare¯ 

Tare out of range 

Tare value exceeds maximum. 

Err 9   Data

 

Internal data error. 

Contact an authorized service agent 

Err13 mMEmM

 

Fail to write EEPROM. 

Contact  an authorized service 
agent 

Err 53 C.SUmM

 

Invalid checksum data 

Contact an authorized service 
agent. 

Lo.Bat 

Battery is discharged. 

Connect the power and charge the 
battery 

Battery fails to charge fully Battery is defective. 

Have battery replaced by OHAUS 
authorized service dealer. 

5.4 Service Information 

 

If the troubleshooting section does not resolve or describe your problem, contact 
your authorized OHAUS service agent.

 

 
6. TECHNICAL DATA 

 

The technical data is valid under the following ambient conditions: 
Indoor use only 
Operating temperature:

 

-10ºC (14ºF) to 40ºC (104ºF) 

Relative humidity: 10% to 90% relative humidity, non-condensing 
Altitude: Up to 2000 m 
Power: AC Adapter (supplied) - 12 VDC 0.84 A output, internal rechargeable sealed 
lead-acid battery 
Mains supply voltage fluctuations: up to ±10% of the nominal voltage 
Installation Category: II 
Pollution Degree: 2 
 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for Valor 2000W Series

Page 1: ...ie Valor 2000W Manual de instrucciones S rie Valor 2000W Manuel d instructions Valor 2000W Serie Bedienungsanleitung Serie Valor 2000W Manuale di istruzioni www GlobalTestSupply com Find Quality Produ...

Page 2: ...conditions specified in these instructions Service should be performed only by authorized personnel Only use weights within the scale s capacity as specified in these instructions Do not operate the s...

Page 3: ...2 Level Indicator 2 5 Power The AC Adapter is used to power the scale when battery power is not needed First connect the AC Adapter plug to the scale input jack located at the bottom of the scale acco...

Page 4: ...battery operation the battery symbol indicates the battery charging status When charging the symbol will blink slowly and when fully charged the symbol will stop blinking TABLE 2 1 Symbol Charge Level...

Page 5: ...ting Short Press Enter clear a Tare value Extended Press3 Initiates Calibration procedure Short Press in Menu Toggles through available settings Notes 1 Short Press Press less than 2 5 seconds 2 Long...

Page 6: ...s displayed 3 3 Calibration Menu To enter the calibration menu extend press Cal button for 5 seconds SpaN will then be displayed Press Zero key to perform span calibration or press Cal key to proceed...

Page 7: ...ow instructions to enter the service menu to set the displayed unit With the scale off and LFT off extend press both the On Zero and Tare keys at the same time for more than 8 seconds until rAmMP is d...

Page 8: ...oggles through active Units Short Press in Menu Toggles through available settings Short Press Initiates an application mode specific response Long Press Selects active Mode Short Press in Menu return...

Page 9: ...nstalled it should be calibrated to ensure accurate results Before performing the calibration be sure to have the appropriate calibration weights as listed in table 3 3 Press and hold Menu until mMeNU...

Page 10: ...the Low limit value Press Yes to accept or No to edit the Low limit value The stored value then displays with the first digit highlighted 000 000 kg Repeatedly press No until the desired number appear...

Page 11: ...cates factory default setting Note The Calibration Menu Mode indicator is displayed when in the Menu Mode 4 6 2 Cal Menu Enter this menu to perform calibrations Span SpaN yes no Initiates a span calib...

Page 12: ...40 05 15 15 14 14 13 13 12 12 11 11 10 40 05 41 12 16 15 15 14 14 13 13 12 12 11 11 41 12 42 19 16 16 15 15 14 14 13 13 12 12 11 42 19 43 26 17 16 16 15 15 14 14 13 13 12 12 43 26 44 32 17 17 16 16 15...

Page 13: ...rtment 5 Remove the plastic pan 6 Clean the plastic pan 6 3 1 2 4 5 Figure 5 1 Pan cleaning 5 2 2 Reinstalling the pan after cleaning Please follow the steps below to reinstall the plastic pan see als...

Page 14: ...of range Tare value exceeds maximum Err 9 Data Internal data error Contact an authorized service agent Err13 mMEmM Fail to write EEPROM Contact an authorized service agent Err 53 C SUmM Invalid checks...

Page 15: ...lay 2 x Red LED front and rear 6 digit 7 segment 20 5 mm 0 8 in characters Keyboard V22PW 2 buttons V22XW 4 buttons Application Modes V22PW Weighing V22XW Weighing Checkweighing Battery Operating Time...

Page 16: ...irective 2006 66 EC introduces new requirements from September 2008 on removability of batteries from waste equipment in EU Member States To comply with this Directive this device has been designed fo...

Page 17: ...standard s requirements On June 21 2012 OHAUS Corporation USA was re registered to the ISO 9001 2008 standard Limited Warranty OHAUS products are warranted against defects in materials and workmanship...

Reviews: