Ranger RC
Interface description
36
5.1.4
Output format
Response to the P command
Field
1 2 3 4 5
6 7 8 9
10
POL SP SP W W W W W W W SP UN UN SP ST SP N N N CR LF
Field 1: POL = polarity, space if positive, - if negative
Fields 2, 4, 6, 8: SP = space
Field 3: W = weight up to 6 digits plus the decimal point
Field 5: UN = unit of measure 1 or 2 characters
Field 7: ST = stability status, space when stable, ? when unstable,
Field 9: N = NET or B/G
Field 10: CR LF = carriage return, line feed
Response to the Print key
Line Field
1 2 3
1
S C A L E : SP SC CR LF
1 2 3 4 5 6 7 8 9
2
G SP SP SP SP SP SP SP SP SP SP SP SP SP POL W W W W W W W SP UN UN SP ST CR LF
3
T SP SP SP SP SP SP SP SP SP SP SP SP SP POL W W W W W W W SP UN UN SP CR LF
4
N SP SP SP SP SP SP SP SP SP SP SP SP SP POL W W W W W W W SP UN UN SP ST CR LF
Line 1
Field 1: SCALE:SP = heading and space
Field 2: SC = scale identifier, 1 = scale 1, 2 = scale 2
Field 3: CR LF = carriage return, line feed
Lines 2, 3, 4
Field 1: G = Gross, T = Tare, N = Net
Field 2: SP = space, up to 13
Field 3: POL = polarity, space if positive, - if negative
Field 4: W = weight, up to 6 digits plus the decimal point
Field 5, 7: SP = space
Field 6: UN = unit of measure, 1 or 2 characters
Field 8: ST = stability status, space when stable, ? when unstable, does not apply to Line 3
Field 9: CR LF =:carriage return, line feed