![OHAUS Ranger RC Instruction Manual Download Page 34](http://html1.mh-extra.com/html/ohaus/ranger-rc/ranger-rc_instruction-manual_1633084034.webp)
Ranger RC
34
Settings in the menu
4.7 Diagnosis and printing out of the menu settings (DIAGNOS)
tESt SC
Internal
Testing scale with internal calibration weight
•
-Int CAL-
appears in the display during the test.
•
After completion of the test, ideally
*d=0.0g
briefly appears in the display, after
which the scale changes to the next menu item
KboArd
.
External
Testing scale with external calibration weight
1. The scale checks the zero point.
-0-
appears in the display. The test weight
flashes in the display.
2. If necessary, change the weight value displayed with
No
.
3. Put the calibration weight on the scale and confirm with
Yes
.
4. The scale checks the calibration weight put on them.
5. After the test is completed, the deviation from the last calibration briefly appears
in the display, ideally
*d=0.0g
, after which the scale changes to the next menu
item
KboArd
.
KboArd
Keyboard test
PUSH 1 ... 10
•
First press the large keys on the bottom row in order:
•
Then press the smaller keys in the top row:
If the key works, the scale changes to the next key.
Note
You cannot abort the keyboard test!
If you have selected the menu item
KboArd
, you must press all keys.
dISPLAY
Display test: The scale displays all functioning segments
SNr
Display of the serial number
SNr2
Display of the serial number of scale 2. This menu item only appears if an analog
second scale is connected.
LiSt
Printout of a list of all menu settings
LiSt2
Printout of a list of all menu settings of scale 2. This menu item only appears if an
analog second scale is connected.
rESEt.AL
Resetting all menu settings to the factory settings
SUrE?
Confirmation inquiry
•
Reset all menu settings to the factory settings with
Yes
•
Do not reset the menu settings with
No
ON/OFF
,
ZERO
,
TARE
,
G/N/T
,
Info
,
Scale Select
,
Count Weight
,
Sample 10
,
Sample Size