EN 7 Ranger
™ 4000 Series
4. Press
Yes
to establish a new reference. Scale will now display [
Put.ref
].
5. Add the desired reference material to the pan or container. Press
Yes
to store the
reference weight. The display shows 100%.
6. Replace the reference material with the sample material. The display shows the
percentage of the sample compared to reference weight.
7. To clear the stored reference press and hold
Mode
until [
Percnt
] is displayed. Press
Yes
when [
Clr.ref
] is displayed.
Note: Press
Function
to view the current reference weight.
3.5 Counting Mode
This mode counts large numbers of items based on the weight of a reference count.
1. Place an empty container on the pan and press
Tare
.
2. Press and hold
Mode
until [
Count
] (Count) is displayed. [
CLr.PwW
] (Clear Average Piece
Weight, APW) will then display.
If no APW exists, the scale will display [
Put.10
], proceed to step 5.
3. Press
No
to use the stored APW. Proceed to step 7.
4. Press
Yes
to establish an APW.
5. The scale will then display the stored sample size, i.e. [
Put 10
]. Press
No
or
Back
to toggle
the choices (5, 10, 20, 50 or 100).
6. Put the indicated number of pieces on the pan then press
Yes
to calculate the APW. The
display shows the piece count.
7. Add additional pieces until the desired count is reached.
8. To clear the stored APW press and hold
Mode
until [
Count
] is displayed. Press
Yes
when
[
Clr.pwW
] is displayed.
Note: Press
Function
to view the current APW.
3.6 Check Mode
Use this mode to compare the Weight, Percent or Count of items to a target weight range. The
scale supports positive, negative and zero check weighing.
Refer to section 4.5 to set the desired Check sub-mode.
3.6.1 Check Weighing
Set Checkweigh to Weight in the Mode menu. Use this mode to compare the weight of items to
a target weight range.
1. Press and hold
Mode
until [
CHeCk
] (Check) is displayed. [
CL.lImM
] (clear check limits) will
then display.
2. Press
No
to use the stored check limits and proceed to step 5.
3. Press
Yes
to establish new check limits. The scale will then display [
Set. Lo
]. Press
Yes
to view the “Low” limit value. Press
Yes
to accept or
No
to edit the “Low” limit value. The
stored value then displays with the first digit highlighted [
0
00.000
kg]. Repeatedly press
No
until the desired number appears. Press
Yes
to accept and highlight the next digit.
Repeat until all the digits are correct. Press
Yes
to accept the “low” limit value, [
Set. Hi
]
will be displayed.
Summary of Contents for Ranger R41MEB
Page 1: ...Ranger 4000 Series Instruction Manual...