Ranger
™ 4000 Series EN 10
4. Manually reset the countdown timer by pressing the
Function
key. [
rEAdY
] (Ready) will
then be displayed.
Note
: If Set0 has been selected, the countdown timer is not displayed. The first weight larger
than 5d will be displayed and held.
Semi-automatic operation
(DYNAMIC is set to SEMI in the Mode menu):
1. Press and hold
Mode
until [
dYNAmM
] (Dynamic) is displayed. [
rEAdY
] (Ready) will then
be displayed.
2. Place the load on the pan to start measurement.
During the average period, the countdown timer decreases in one second increments.
3. After completed countdown, the average weight will be displayed. The tilde symbol will
blink indicating that the current weight is being held.
4. Manually reset the countdown timer by pressing the
Function
key. [
rEAdY
] (Ready) will
then be displayed.
Note
: If Set0 has been selected, the countdown timer is not displayed. The first weight larger
than 5d will be displayed and held.
Automatic operation
(DYNAMIC is set to AUTOMATIC in the Mode menu):
1. Press and hold
Mode
until [
dYNAmM
] (Dynamic) is displayed. [
rEAdY
] (Ready) will then
be displayed.
2. Place the load on the pan to start measurement.
During the average period, the countdown timer decreases in one second increments.
3. After completed countdown, the average weight will be displayed. The tilde symbol will
blink indicating that the current weight is being held.
4. Remove the load and the display will reset after 10 seconds, [
rEAdY
] (Ready) will then be
displayed.
Notes
: If Set0 has been selected, the countdown timer is not displayed. The first weight larger
than 5d will be displayed and held.
Alternatively, the display may be manually reset by pressing the
Function
key.
3.8 Accumulation and Statistics
The Accumulation feature enables manual or automatic totalizing of displayed values. Statistical
data is stored in memory for review and printing. Accumulation works together with each
application mode except Dynamic.
3.8.1 Accumulating Displayed Values
With ACCUMULATE set to MANUAL, place the item on the scale and press the
M+
key to add
the weight to
accumulation data. The ∑ icon will keep flashing until the weight is removed.
With ACCUMULATE set to AUTO, place the item on the scale. The displayed value is
accumulated automatically.
The ∑ icon will keep flashing until the weight is removed.
3.8.2 Viewing and Clearing Statistical Data
When the pan is cleared, press the
M+
key to
display the statistical information. To clear the accumulation data press the
Tare
key while the
statistical information is displayed. The display shows [
Clr.aCC
]. Press the
Yes
key to clear the
stored data and return to current mode.
Summary of Contents for Ranger R41MEB
Page 1: ...Ranger 4000 Series Instruction Manual...
Page 2: ......