background image

EN 13                                                                                                        Ranger

™ 4000 Series 

 
For menu items with numeric settings such as Capacity, the current 
setting is displayed with all digits flashing. 
Press the 

No

 button to begin editing. 

The first digit is displayed flashing. 
 
 
Press the 

No

 button to increment the digit or press the 

Yes

 button to 

accept the digit and move to the next digit. 
 

Repeat this process for all digits. 

 

Press the 

Yes

 button when the last digit has been set.  

 

The new setting is displayed with all digits flashing. Press the 

Yes

 

button to accept the setting or press the 

No

 button to resume editing. 

This method also applies to setting Checkweigh under and over 
targets. 

For End menu items, pressing the 

Yes

 button advances to the next 

menu, while pressing the 

No

 button returns to the top of the current 

menu.

 

 

4.2 Calibration Menu

 

Enter this menu to perform calibrations.

 

 
Initial Calibration 

When the scale is operated for the first time, a span calibration is recommended to ensure 
accurate weighing results. Before performing the calibration, be sure to have the appropriate 
calibration weights as listed in table 4-1. Ensure that the LFT switch/calibration lock is set to 
unlocked position. See figure 5-1.  

Or adjust the GEO setting according to your location (see table 4-2). 

Procedure: 

Press and hold 

Menu

 until [

mMeNU

] (Menu) is displayed. When the button is released, the display 

will show [

C.A.L

]. Press 

Yes

 to accept. [

SpaN

] will then be shown. Press 

Yes

 to begin the span 

calibration. [

   0

 

kg

] will be displayed. Press 

Yes

 to accept. [

--C--

] will be displayed while zero 

reading is stored. Next, the display shows the calibration weight value. Place the specified 
calibration mass on the pan. Press 

Yes

 to accept the weight or 

No

 to select an alternate weight. 

[

--C--

] will be displayed while the reading is stored. The display will show [

done

] if the calibration 

was successful. The scale returns to the previous application mode and is ready for use.  

 

TABLE 4-1 

Required Span Calibration Mass (sold separately) 

Max 

Mass

1

 

Max 

Mass

1

 

3000g 

3kg / 6lb 

15000g 

15kg / 30lb 

6000g 

6kg / 15lb 

30000g 

30kg / 60lb 

                
 Note:

 

1

 Pound masses are used when calibrating in the lb unit. 

Summary of Contents for Ranger R41MEB

Page 1: ...Ranger 4000 Series Instruction Manual...

Page 2: ......

Page 3: ...e safety precautions Verify that the AC input voltage printed on the data label matches the local AC power supply Only connect the power cord to a compatible grounded electrical outlet Do not position...

Page 4: ...s heat sources or rapid temperature changes Allow sufficient space around the scale 2 4 Leveling the Equipment The Ranger Series has a level indicator as a reminder that the scale should be leveled fo...

Page 5: ...in table 3 2 will remain lit The scale can be operated during charging and the battery is protected against overcharging For maximum operating time the battery should be charged at room temperature D...

Page 6: ...s the accumulation when the accumulation information is displayed Secondary Function Long Press Off Turns the scale Off Units Changes the weighing unit Mode Allows changing the application mode Menu E...

Page 7: ...l 11 Pound Ounce Pound Ounce symbols 5 Preset Tare Tare symbols 12 Kilogram gram symbols 6 Pointer symbols 13 Pieces symbol 7 Accumulation symbol 14 Percent symbol The colored LED indicators on the le...

Page 8: ...gh is displayed 2 If required place an empty container on the pan and press Tare 3 Add sample to the pan or container The display shows the weight of the sample 3 4 Percent Mode This mode measures the...

Page 9: ...d number of pieces on the pan then press Yes to calculate the APW The display shows the piece count 7 Add additional pieces until the desired count is reached 8 To clear the stored APW press and hold...

Page 10: ...hen display Set Lo Press Yes to view the Low limit value Press Yes to accept or No to edit the Low limit value The stored value then displays with the first digit highlighted 000 000 Repeatedly press...

Page 11: ...han the UNDER limit Add material to the scale until it is within the ACCEPT green range Negative Check Negative check is used to determine when the material removed from the scale is within the target...

Page 12: ...ncrements 3 After completed countdown the average weight will be displayed The tilde symbol will blink indicating that the current weight is being held 4 Remove the load and the display will reset aft...

Page 13: ...ting select mMaNU If required place an empty container on the pan and press Tare Step i Put the weight 0 04 kg on the pan and press M key The icon indicator will keep flashing until the weight is remo...

Page 14: ...vailable in all models 1 This menu item is only used in models with LED display When LEGAL FOR TRADE is set to ON Lock Switch in locked position the menu settings are affected as follows Calibration C...

Page 15: ...accurate weighing results Before performing the calibration be sure to have the appropriate calibration weights as listed in table 4 1 Ensure that the LFT switch calibration lock is set to unlocked po...

Page 16: ...the top of the current menu 4 3 Setup Menu Enter this menu to set scale parameters Reset reset Reset the Setup menu to factory defaults NO not reset YES reset Power on unit pwWr UN Set the unit of me...

Page 17: ...UNDER the beeper will sound when the weight is above the Over setting OVER the beeper will sound when the weight is below the Under setting or above the Over setting Beeper Key BP key Set whether the...

Page 18: ...the display will maintain zero until a change of 1 divisions per second has been exceeded 3d the display will maintain zero until a change of 3 divisions per second has been exceeded Light LIGHt LCD...

Page 19: ...t Set the status OFF disabled ON enabled Count COUNt Set the status OFF disabled ON enabled Check CHECk Set the sub mode OFF disabled WEIGH weight mode PERCNT percent mode COUTN count mode Dynamic DYN...

Page 20: ...le Only Stable Set the printing criteria OFF values are printed immediately ON values are only printed when the stability criteria are met Auto Print A Print Set the automatic printing functionality O...

Page 21: ...he printed data Result Set the status OFF disabled ON the displayed reading is printed Gross Set the status OFF disabled ON the gross weight is printed Net Set the status OFF disabled ON the net weigh...

Page 22: ...ows XP Setup 1 Click Start Menu in Windows XP and click Settings open Control Panel 2 Double click Accessibility Options in Control Panel 3 Select the General tab 4 Check Use Serial Keys and click the...

Page 23: ...ODD 7 data bits odd parity 7 NONE 7 data bits no parity 8 NONE 8 data bits no parity Stop bit stOP Set the number of stop bits 1 1 stop bits 2 2 stop bits Handshake H shake Set the flow control metho...

Page 24: ...tion which can lock menu settings to prevent tampering Reset reset Reset the Lock menu to factory defaults NO no reset YES reset Lock Cal l CAL Set the status OFF Calibration menu is not locked ON Cal...

Page 25: ...nu Press Yes to advance to the Calibration menu Press No to exit the menu and return to the current application mode 4 11 Additional Features The Weigh Below Hook is provided with the scale To use thi...

Page 26: ...1 42 19 43 26 17 16 16 15 15 14 14 13 13 12 12 43 26 44 32 17 17 16 16 15 15 14 14 13 13 12 44 32 45 38 18 17 17 16 16 15 15 14 14 13 13 45 38 46 45 18 18 17 17 16 16 15 15 14 14 13 46 45 47 51 19 18...

Page 27: ...ned in Section 4 2 3 Set the switch to Locked See figure 5 2 The Menu Lock switch limits changes to the Cal Setup Readout Unit and Print menus The switch in type approved models may set some scale set...

Page 28: ...Ranger 4000 Series EN 26 Figure 5 2 Sealing Paper Seal Wire Seal Unlocked Locked with Paper Seal Locked with Wire Seal Lock Switch...

Page 29: ...abled Enter menu and enable unit Battery icon flashing Low Battery error Connect scale to AC power and charge the battery Err 8 1 Power On Error Weight reading exceeds Power On Zero limit Err 8 2 Powe...

Page 30: ...tween 8 00 AM and 5 00 PM EST An OHAUS product service specialist will be available to provide assistance Outside the USA please visit our web site www ohaus com to locate the OHAUS office nearest you...

Page 31: ...00 g x 1 g 6 lb x 0 002 lb 96 oz x 0 04 oz 6 kg x 0 002 kg 6000 g x 2 g 15 lb x 0 005 lb 240 oz x 0 1 oz 15 kg x 0 005 kg 15000 g x 5 g 30 lb x 0 01 lb 480 oz x 0 18 oz 30 kg x 0 01 kg 30000 g x 10 g...

Page 32: ...Print of displayed weight stable or unstable P Print displayed weight stable or unstable CP Continuous Print SP Print on Stability xS 0S Turn off Stable Only menu item and allow unstable print 1S Tur...

Page 33: ...mands I1 Inquiry of SICS level and SICS versions I2 Inquiry of scale data I3 Inquiry of scale software version I4 Inquiry of serial number S Send stable weight value SI Send weight value immediately S...

Page 34: ...s entered 1 23 kg T Tare value line If Printx Content Tare is ON and a tare value is entered REF WGT 0 012 kg Information line If Printx Content Info is ON MODE PERCENT Mode line If Printx Content App...

Page 35: ...e left of most significant character if negative Unit Up to 5 characters left justified If the Unit in the Print Content menu was set to OFF the unit will be removed in the weight string and replaced...

Page 36: ...e manufacturers work It comprises all tests according to the adopted European standard EN 45501 1992 paragraph 8 2 2 If national regulations limit the validity period of the verification the user of t...

Page 37: ...can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Inc...

Page 38: ...been damaged by accident or misuse exposed to radioactive or corrosive materials has foreign material penetrating to the inside of the product or as a result of service or modification by other than...

Page 39: ...or del mundo Avec des bureaux dans le monde entier Weltweite Gesh ftsstellen Con uffici in tutto il mondo www ohaus com 30240427 P N 30240427 2020 OHAUS Corporation all rights reserved todos los derec...

Reviews: