![OHAUS Pioneer PJX Series Service Manual Download Page 49](http://html.mh-extra.com/html/ohaus/pioneer-pjx-series/pioneer-pjx-series_service-manual_726914049.webp)
APPENDIX C – SERVICE TOOL
Pioneer PJX Series Service Manual
43
Ohaus Corporation www.ohaus.com
APPENDIX C
–
SERVICE TOOL
The Software Service Tool (Part Number 83032124) is required when a main PC Board is replaced in
a Pioneer PJX Series balance. It is used to re-configure the balance to its original parameters in the
case of a PCB replacement. The tool can also be used to communicate with the balance using
commands that are listed at the end of the appendix.
The latest software service tool and support files are available on the Ohaus DMX site.
Please read the Service Tool Instruction Manual (Part Number 30032352) which is also available on
Ohaus DMX site.
C.1
BALANCE COMMANDS
The balance commands are used to send instructions to the balance. Commands listed in the
following table will be acknowledged by the balance. The balance will return “ES” for invalid
commands.
TABLE C-1
INTERFACE COMMAND LIST
Command
Function
IP
Immediate Print of displayed weight (stable or unstable).
P
Print displayed weight (stable or unstable).
CP
Continuous Print.
SP
Print on Stability.
SLP
Set Auto Print to On Stability, allow non-zero displayed weight be printed.
SLZP
Set Auto Print to On Stability, allow both stable non-zero weight and stable zero reading
printed.
xP
Set Auto Print to Interval Print, x = print interval (1-3600 sec), 0P disable the interval Print
0P
0P disable interval print, continuous print or print on stability
H
Enter Print Header Lines, the format is: H x "header string".
Where x = line number 1 to 3, “header string” can be up to 24 alphanumeric characters.
If no string in the command, "H x" will read the stored header x.
Z
Same as pressing Zero Key.
T
Same as pressing Tare Key.
xT
Establish a preset Tare value in displayed unit. x = preset tare value. Sending 0T clears tare
(if allowed).
PT
Prints Tare weight stored in memory.
PM
Print current application mode (weighing mode).
xM
Set current application mode to x. x depends on applications
1 – Weigh
2 – Count
3 – Percent
4 – Density
M
Scroll to the next enabled mode.
PU
Print Current weighing unit: g, kg, lb, oz, etc….
xU
Set balance to unit x: g, kg etc.
1 – g
2 – kg
3 – mg
4 – ct
…
Summary of Contents for Pioneer PJX Series
Page 1: ...Pioneer PJX Series Balances Service Manual ...
Page 2: ......
Page 6: ......