APPENDIX B – SERVICE MENU
Ohaus Corporation www.ohaus.com
40
Pioneer PJX Series Service Manual
APPENDIX B - SERVICE MENU
This appendix describes Service Menu, and entry by following step:
– Turn the balance off
– Press and Hold the
Zero
button and the
Tare
/Menu-Cal Button. As the balance powers up,
Service
appears followed by
Ramp
Service Sub-Menu
RAMP
The ramp display shows the percentage of use of the A to D circuit, that is, of the temperature-
compensated duty cycle. The actual value is not as important as how it changes. It should increase
as the weight on the balance is increased. The ramp display should remain constant without
fluctuations.
The normal range is within 3% to 97%. If the readings are outside this range the load-cell is likely
bad. Fluctuations in the display may indicate a mechanical interference, a cable connection
problem, a damaged Main PCB or a damaged load-cell.
SERVICE SPAN CALIBRATION
Service span calibration allows a new zero and maximum setting using an external calibration
weight.