APPENDIX A - USER CALIBRATIONS
Pioneer PJX Series Service Manual
39
Ohaus Corporation www.ohaus.com
3. To change the calibration mode, press
No
until
LineRr
(linearity calibration) is displayed.
4. The calibration mass value will be shown in the display.
After the display shows
100.0000
g, place weight(s) of
100 g on the pan for calibration.
5. Remove the weight(s) of 100 g from the pan. After a
while,
200.0000
g will be displayed on the screen. Place
weight(s) of 200 g on the pan.
6. Once the linearity calibration is completed successfully,
CAL.done
will display.
1.2
Press any button to return to the previous
screen.
7. Remove the weight(s), and the reading will be set to
zero.