CHAPTER 4 - TESTING
Pioneer PJX Series Service Manual
25
Ohaus Corporation www.ohaus.com
4
TESTING
Before and after servicing an Pioneer PJX Series balance, an operational test and various
performance tests should be made to confirm if the balance meets specifications. Turn the balance on
and allow it to warm up for at least one hour before performing these tests. The Analytical models
need up to 4 hours to stabilize.
NOTE:
Make sure the test area is free from drafts and that the balance rests on a level and vibration-
free surface. The Analytical models especially need a solid platform, such as a stone table.
4.1
TEST MASSES REQUIRED
The masses required to test the Ohaus Pioneer PJX Series balances must meet the requirements
of the ASTM or OIML Tolerances listed in the table. Poor quality calibration masses can be the
cause of frustrating diagnostics.
TABLE 4-1
CALIBRATION MASS VALUES
InCal Model
PJX
ExCal
Model
PJX
Span
Calibration
Points (g)
Linearity Calibration Points
(g)
Weight Class
PJX203(M)
PJX203/E
20, 40
10, 20, 30, 40
ASTM Class 1
OIML E2
PJX803(M)
PJX803/E
80, 160
40, 80, 120, 160
ASTM Class 1
OIML E2
PJX1203(M)
PJX1203/E
100, 200
50, 100, 150, 200
ASTM Class 1
OIML E2
PJX1603(M)
PJX1603/E
150, 300
100, 200, 300
ASTM Class 2
OIML F1
PJX2503(M)
PJX2503/E
250, 500
200, 300, 400, 500
ASTM Class 2
OIML F1
PJX822(M)
PJX822/E
400, 800
200, 400, 600, 800
ASTM Class 3
OIML F2
PJX1602(M)
PJX1602/E
750, 1500
500, 1000, 1500
ASTM Class 3
OIML F2
PJX2202(M)
PJX2202/E
1000, 2000
500, 1000, 1500, 2000
ASTM Class 3
OIML F2
PJX4202(M)
PJX4202/E
2000, 4000
1000, 2000, 3000, 4000
ASTM Class 2
OIML F1
PJX5202(M)
PJX5202/E
2500, 5000
2000, 3000, 4000, 5000
ASTM Class 2
OIML F1
PJX2201
PJX2201/E
1000, 2000
500, 1000, 1500, 2000
ASTM Class 4
OIML F2
PJX4201(M)
PJX4201/E
2000, 4000
1000, 2000, 3000, 4000
ASTM Class 4
OIML F2
PJX6201
PJX6201/E
3000, 6000
2000, 4000, 6000
ASTM Class 4
OIML F2
4.2
OPERATIONAL TEST
1. Connect a functioning Power Adapter to the balance.
2. Plug the Power Cord into a suitable power source.
3. Verify that the balance start up sequence occurs properly. Note any error codes.
4.3
LOAD CELL TEST USING RAMP
To test the Load Cell using RAMP, see
Appendix B
.