APPENDIX A - USER CALIBRATIONS
Pioneer PJX Series Service Manual
37
Ohaus Corporation www.ohaus.com
A.2
SPAN CALIBRATION
Span calibration uses two calibration points, one at
zero load
and the other at
specified
full load
(span). For detailed calibration mass information please see the specification tables 4-1.
With the balance turned On and no load on the pan, Span Calibration can be performed. The best
accuracy is achieved using the mass closest to the full span value.
Steps for span calibration
1. Press and hold the
Tare / Menu-Cal
button, and the
Calibration Menu
will display.
2. Press
Yes
to enter the Calibration Menu.
3. To change the calibration mode, press
No
until
SpaN
(span calibration) is displayed.
4. The Calibration mass value will be shown in the screen.
After the display shows
200.0000
g, place weight(s) of
200 g on the pan for calibration. To change to the
calibration point of half full capacity (e.g. 100 g), press
the Function / Mode button. After the screen shows
"Place weights" and "100.000 g", place weight(s) of 100
g on the pan for calibration.