CHAPTER 1 GETTING STARTED
Ohaus Corporation www.ohaus.com
1-8
Pioneer
™ Balances Service Manual
1.5.7 The Load Cell
Figure 1-4 illustrates a typical Magnetic Force Restoration (MFR) Load Cell.
The force of a weight placed on the Weighing Pan moves the Hanger which is guided by the
Upper and Lower Flexure Arms. The Load Flexure transfers the force vertically to the Ratio
Beam. The Ratio Beam Flexures form a pivot for the Ratio Beam. When the Ratio Beam is
moved, the Force Coils mounted on the Ratio Beam are moved. The Force Coils are located in
the magnetic field created by the Permanent Magnets. The magnetic forces are affected by
temperature. A Temperature Sensor is attached near the magnets to measure the temperature
and allow the electronics to compensate for the changes in temperature.
Operation
A mass placed on the Pan applies a downward force to the Load Flexure by means of the
Hanger. The Load Flexure moves one end of the Ratio Beam which pivots, moving the Force
Coil up into the air gap of the Magnet. The Force Coils are located in the field of the Permanent
Magnets. The movement of the Vane (which is connected to the Ratio Beam) is detected by the
Position Sensor. The Position Sensor contains a sender and receiver. When the Vane moves,
the amount of light sent to the receiver also changes. The change in the position of the Sensor
Vane is measured and used along with the Temperature Sensor signal to adjust the current in
the Force Coils. The current in the Force Coils produces a magnetic field which restores the
Ratio Beam to its normal position.
Figure 1-4. Load Cell Assembly.
Ratio Beam
Flexure
Vertical
Flexure
Top Bracket
Hanger
Bottom Flexure Arm
Flexure
Magnet
Printed Circuit
Board
Temperature
Sensor
Position
Sensor
Vane
Force
Coils
Ratio
Beam
Counter
Weight
Upper Flexure Arm
Pan
Summary of Contents for Pioneer PA64
Page 2: ......
Page 10: ...TABLE OF CONTENTS Pioneer Balances Service Manual viii Ohaus Corporation www ohaus com ...
Page 115: ......