CHAPTER 3 MAINTENANCE PROCEDURES
Ohaus Corporation www.ohaus.com
3-32
Pioneer Balance Service Manual
3.3.18 Reassembling Analytical Load Cells
17. Install the top bracket. (See Figures 3-92 and 3-93.)
Figure 3-92. Analytical InCal Load Cell
with Top Bracket visible.
Figure 3-93. Analytical balance Load Cell
with Top Bracket visible.
18. Remove the screws and shims
from the Hanger.
19. Install the Bottom Mounting Plate
as shown. Note the position of the
Weigh Below Hook. (See Figure
3-94.)
Figure 3-94. Bottom Mounting Plate on
Analytical Load Cell.
20. Position the Load Cell in the
balance Housing. Insert and
tighten the three screws, from
underneath the Bottom Housing.
Be sure the Load Cell is squarely
mounted in the Housing. Connect
the cable from the Main PCB to the
Position Sensor Board. (See
Figure 3-95.)
Figure 3-95. Ribbon cable connects PSB to
Main PCB.
Cable from Main
PCB to Position
Sensor Board
Top Bracket
Top Bracket (with
InCal Weight Arms)
Screws on
Bottom Mounting
Plate
Screws that secure
the Flexure Arms
to the Load Cell