3.3.8 Load Cell Disassembly
– Precision Balances
18. Note the position of the opening and then turn the black Up/Down Stop Screw, so the
Ratio Beam can clear the stop, and gently lift out the Ratio Beam.
Figure 3-31. Ratio Beam clearing stop on
Up/Down Stop Screw as it is lifted out.
Figure 3-32. Magnets must be clean.
The Load Cell is now completely disassembled.
To reassemble the Load Cell, see Section 3.3.11, Load Cell Reassembly
– Precision Balances.