CHAPTER 3 MAINTENANCE PROCEDURES
Pioneer
™ Balance Service Manual
3-3
Ohaus Corporation www.ohaus.com
3.3 LOAD CELL REMOVAL AND DISASSEMBLY
The Load Cell may need to be removed and repaired because of balance instability, or because
the balance does not calibrate or repeat, or because the balance is physically broken or
displays an error code. The ratio coil may have a short circuit. Possible causes of Load Cell
problems include: parts being misaligned, bent or broken Flexures, contamination or debris, or
in any moving part.
3.3.1 Variations in Pioneer Load Cells
Pioneer balances have different types of Load Cell. Use the illustrations below to identify the
type used in the balance being serviced.
Figure 3-3. Precision Load Cell with InCal:
Section 3.3.2, p. 3-5
.
Figure 3-4. Analytical Load Cell with InCal:
Section 3.3.6, page 3-7
.
Figure 3-5. Precision Load Cell:
Section 3.3.6, page 3-7
.
Figure 3-6. Analytical Load Cell:
Section 3.3.14, page 3-20
.
For Block Style Load Cells, see next page
Precision & Analytical Load Cells have different style Flexure Arms.