APPENDIX A CALIBRATION
MB23 and MB25 Service Manual
A-3
Ohaus Corporation www.ohaus.com
A.3 SERVICE CALIBRATION
Span calibration from the service menu allows you to set a new zero and maximum setting.
This is distinct from user level span calibration, which allows a user to adjust the zero and
maximum setting within the range established by the service menu span setting.
Remove any load from the Pan, including sample pan. Analyzer must be
OFF but plugged in.
Have 100g and 50g calibration masses available.
Note
: To cancel calibration without saving changes, press:
6
(Release keys...)
1
(Place weight on pan.)
SErv
SPAN
100.00
g
(Clear pan.)
----
0
g
----
-done-
1
(Place weight on pan.)
lINea
100.00
g
(Clear pan.)
----
0
g
(Place weight on pan.)
(Place weight on pan.)
50.00
g
100.00
g
(Place weight on pan.)
(Clear pan.)
50.00
g
0
g
(Place weight on pan.)
(Clear pan.)
100.00
g
0
g
done
end
1
Off