APPENDIX A STANDARD CALIBRATION
Ohaus Corporation www.ohaus.com
A-2
MB23 and MB25 Service Manual
A.2 TEMPERATURE CALIBRATION
Use a Temperature Calibration Kit. (Available as an accessory.) If the analyzer
has been recently used, allow at least 30 minutes before performing
calibration.
Use
only
the OHAUS Temperature Calibration Kit. Using other
methods could damage the Analyzer.
1
(Remove pan and pan support.)
(Appears in software prior to 1.07.)
0.000
g
Err8.4
3
CAL
tc100
15
(Beep & Blink)
Match:
/
100
106
1
15
(Beep & Blink)
Read Temp
thru Grating
Match:
/
tc160
160
166
1
Open cover:
done
Analyzer returns to Weigh mode.
Allow Temperature Calibration Unit to cool before removing it.
Read Temp thru
Grating