EN-32
Defender 3000 Indicators
Grad2
[
GrAd2
]
Set the second readability of the scale.
Note
:
The value of graduation 2 must be larger than graduation 1, and it applies to capacity as well.
The capacity and graduation value should be within the following range, or your setting will be
unsuccessful:
Capacity / 30000 ≤ Graduation ≤ Capacity / 600
Power On Zero
[
P.ZerO
]
Zero the scale at Power On.
Off
= disabled.
ON
= enabled.
Power On Unit
[
P.UNIt
]
Set the unit that will be displayed at Power On.
AUtO
= last unit in use when turned off
kg
= kilograms
g
= grams
lb
= pounds
oz
= ounces
lb:oz
= pound ounces
t
= metric tonne
Auto Tare
[
A.
tArE
]
Set the automatic tare feature.
Off
= automatic tare is disabled.
ON
= the first stable gross weight is tared. The tared value will be cleared when the indicator returns
to gross zero.
ACCept
= stable gross weights within the Accept limits are tared (in Check mode only).