Defender 3000 Indicators
EN-31
4.3
Setup Menu
Enter this menu
S.E.t.U.P
to set scale parameters. Default settings are in
bold
.
Menu
Sub-Menu
Sub-Menu (in
segment)
Options
Options (in segment)
Setup
S.E.t.U.P
Reset
reset
no
, yes
NO, yes
Capacity Unit
C.UNIt
kg
, lb
/
Range
rANGE
Single
, Dual
SINGLE
,
DUAL
>|1|<
Capacity
CAP
0.1-99999.9
/
>|1|<
Graduation
Grad
0.0001~100
/
>|2|<
Capacity
CAP
0.1-99999.9
/
>|2|<
Graduation
Grad
0.0001~100
/
Power On Zero
P.ZerO
Off,
On
Off
,
ON
Power On Unit
P.UNIt
Auto
, g, kg, lb, oz, lb:oz, t
AUtO
Auto Tare
A.tare
Off
, On, Accept
Off
,
ON
,
ACCept
End
End
/
/
Note
: / refers to not applicable in the table above.
Reset
[
reset
]
Reset the Setup menu to factory defaults.
NO
= do not reset
yes
= reset
Capacity Unit [
C.
UNIt
]
Select the unit used for calibration.
kg
lb
Range
[
rANGE
]
Selects the number of weighing intervals.
This terminals can be configured to use single or dual interval. Each interval can be assigned its own
graduation. If dual interval is selected, the graduation will change when the weight reaches the second
interval.
When
Single
interval is selected, the additional parameters available are:
Capacity
Graduation
When
Dual
interval is selected, the terminal functions with two intervals, each with its own capacity and
graduation. In addition to the Interval 1 capacity and graduation parameters, the following two parameters
are available:
Capacity 2
Graduation 2
Capacity
[
Cap
] /
Capacity1
[
Cap1
]
Set the capacity of the scale or the first capacity of the scale.
0.1-99999.9
Grad
[
GrAd
] /
Grad1
[
GrAd1
]
Set the scale readability or the first readability of the scale.
0.0001~100
Capacity2
[
Cap2
]
Set the second capacity of the scale.