OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
14
2. The bath should be checked and cleaned annually.
a. Access the bath housing as described in steps a through f on the
previous page.
b. Visually inspect the mineral oil in the bath. If it contains any foreign
material, pour out the entire contents and refill with clean mineral oil.
3. Slurry containers and paddles should be kept as clean as possible. The
lid should be thoroughly cleaned and any build up of cement should be
removed.
4. The unit has four fuses, located on the front panel beneath the four power
switches. If any of the switches stop working, check the corresponding
fuse. To remove a fuse, twist and pull the knob. After replacing the fuse,
reinsert it into the slot and twist it to secure it in place.
5. To replace the main bearings:
a. Remove all of the set screws in the front and rear side of the drive
housing.
b. Push the bearing up and out from below.