5
Maintenance
The pump must only be opened by such persons who have been
authorised by Oerlikon Leybold Vacuum to do so.
5.1
Maintenance Intervals
After 15,000 operating hours at the latest a standard bearing exchange will
be recommended for the
TURBOVAC 35
to
361 C
. For the
TURBOVAC 600 C
and 1000 C
a standard bearing exchange will be required after 10,000 opera-
ting hours at the latest.
Moreover we recommend to have the rotor assy. exchanged , depending on
the thermal stress on the rotor, after 45,000 to 100,000 operating hours. This
can only be done by Oerlikon Leybold Vacuum Service. For this ask for a
quotation.
When using purge gas valves
Depending on the degree of contamination of the purge gas used the filter
will clog and will have to be exchanged (our experience indicates that this will
become necessary after 1 to 6 months).
When using an adsorption trap
Regenerate or renew the adsorption agent regularly; refer to the operating
instructions provided with the trap.
5.2
Oerlikon Leybold Vacuum Service
Whenever you send us in equipment, indicate whether the equipment is con-
taminated or is free of substances which could pose a health hazard. If it is
contaminated, specify exactly which substances are involved. You must use
the form we have prepared for this purpose.
A copy of the form has been reproduced at the end of these Operating
Instructions: “Declaration of Contamination for Compressors, Vacuum Pumps
and Components”. Another suitable form is available from www.oerlikon.com
→
Oerlikon Leybold Vacuum Systems
→
Documentation
→
Download
Documents.
Attach the form to the equipment or enclose it with the equipment.
This statement detailing the type of contamination is required to satisfy legal
requirements and for the protection of our employees.
We must return to the sender any equipment which is not accompanied by a
contamination statement.
Maintenance
29
GA05118_1502 - 12/2006 - © Oerlikon Leybold Vacuum
Contamination
Form
Caution
Standard bearing exchange