Water cooling
When attaching the water cooling unit to the
TURBOVAC
35 LS, 35 LS2 and
50, remove the pump foot and then bolt the cooling unit to the bottom of the
pump. The mounting bolts are provided with the water cooling unit.
Connect the coolant hoses to the hose nipples and secure with hose clamps.
If the coolant flow is turned on and off by means of a solenoid valve, make
the electrical connection in such a way that coolant flow will be started and
stopped together with the pump itself.
Cooling water specifications
Inlet pressure
3 to 7 bar absolute
Cooling water requirement, inlet temperature
See Fig. 7
Appearance
Colorless, clear, free of oils and greases
Sediments
< 250 mg/l
Particle size
< 150 µm
pH value
7 to 8.5
Overall hardness (total alkaline earths)
max. 20
°
German hardness scale
(= 3.57 mmol/l)
Installation
21
GA05118_1502 - 12/2006 - © Oerlikon Leybold Vacuum
5
10
15
20
25
30
35
°
C
50
40
30
20
10
0
l/h
°
C
50
40
30
20
10
0
l/h
5
10
15
20
25
30
35
°
C
80
70
60
50
40
30
20
10
0
l/h
5
10
15
20
25
30
35
Fig. 7 Cooling water consumption
Cooling water temperature
Cooling water flow
TURBOVAC 35 LS, 35 LS2, 50
TURBOVAC 600, 1000
TURBOVAC 151, 361
Cooling water flow
Cooling water flow
Cooling water temperature
Cooling water temperature