C: Analog Output
013-20000-009-02-201510
51
Analog Data Range Example
As an example, consider the following. A user who is interested in using the analog voltage output to
monitor oxygen in the range from 4.5 ppm to 7 ppm would select the following settings. The result would
be that any ppm values below 4.5 ppm would result in an output of 0 volts and any values above 7 ppm
would result in 5 volts. In this case, 6.3 ppm would produce an analog voltage output of [(6.3 - 4.5) / (7 -
4.5)] * 5 = 3.6 volts.
Analog Voltage Source: 5 – Alternate O2 Units
Analog Voltage Lower Bound: 4.5
Analog Voltage Upper Bound: 7.00
(Oxygen Units) – “DO – ppm – Water”. (Selection of ppm as alternate O2 units)
Saving the Analog Output Settings
To save the analog output settings, open the “Advanced Settings” control panel and then select “Enabled”
from the dropdown box located to the right of the “Flash Write” setting. This box will momentarily
continue to display “enabled”, but will switch back to “disabled” when the unit has processed the
command.
Analog Output Pinouts
Caution
Be careful when wiring. Incorrect wiring could cause permanent damage to the unit.
View of Connector on Back of NeoFox Unit
1 = RS232 Rx
2 = 0-5V
– V+
3 = Current 4-20 - V- (
Iout
)
4 = Current 4-20 - V+
5 = RS232 Tx
6 = 0-5V / RS232 GND
This is the view looking at the connector on the back of the NeoFox, with power plug on the right
and the USB plug on the left. The connector is a standard PS-2 connector.
Analog Output Parameters List
NeoFox Viewer Parameter
Potential Values
Voltage Output Source
0
– Disable analog output
1
– Oxygen
2
– Temperature (not supported)
3
– Pressure (not supported)
4
– Tau
Summary of Contents for NeoFox
Page 22: ...2 Installation 14 013 20000 009 02 201510 ...
Page 42: ...4 Using NeoFox Sport 34 013 20000 009 02 201510 NeoFox Sport Menu Menu Tree Description ...
Page 48: ...4 Using NeoFox Sport 40 013 20000 009 02 201510 ...
Page 50: ...A Specifications 42 013 20000 009 02 201510 ...
Page 54: ...B Maintenance 46 013 20000 009 02 201510 ...
Page 63: ...C Analog Output 013 20000 009 02 201510 55 ...
Page 64: ...C Analog Output 56 013 20000 009 02 201510 ...
Page 66: ...D Calculating Drift Per Sensor 58 013 20000 009 02 201510 ...