3: Using NeoFox with NeoFox Viewer Software
013-20000-009-02-201510
23
2. Click the
button in the
Gain
field.
3. Slide the APD Voltage selector to the desired level.
4. Click the
button in the
Gain
field.
Calibrating the Probe
When you access the software for the first time, the Oxygen field will be blank. To see an oxygen value,
you must first enter calibration information for the probe. This calibration information is unique for each
oxygen sensor probe itself, but is independent of the NeoFox hardware unit. Therefore, a new calibration
must be loaded each time a new sensor is added or before first use.
You have the option of purchasing a factory calibration for your Oxygen Sensor Probe (FOXY-CAL,
HIOXY-CAL, FOSPOR-CAL). If you have a factory calibration, follow the instructions for
. If you do not have a factory calibration, follow the instructions for a
Multipoint Calibration
Follow these instructions if you purchased a factory calibration file for your sensor. This file can be found
on the NeoFox CD. To retrieve it, select Calibration from the Options menu.
►
Procedure
1. From the Main screen, select
Options | Calibration
.
Summary of Contents for NeoFox
Page 22: ...2 Installation 14 013 20000 009 02 201510 ...
Page 42: ...4 Using NeoFox Sport 34 013 20000 009 02 201510 NeoFox Sport Menu Menu Tree Description ...
Page 48: ...4 Using NeoFox Sport 40 013 20000 009 02 201510 ...
Page 50: ...A Specifications 42 013 20000 009 02 201510 ...
Page 54: ...B Maintenance 46 013 20000 009 02 201510 ...
Page 63: ...C Analog Output 013 20000 009 02 201510 55 ...
Page 64: ...C Analog Output 56 013 20000 009 02 201510 ...
Page 66: ...D Calculating Drift Per Sensor 58 013 20000 009 02 201510 ...