3: Using NeoFox with NeoFox Viewer Software
26
013-20000-009-02-201510
3. Under Select Sensor Calibration Method, choose Two-Point Calibration.
4. Under Select How Temperature Will Be Measured, choose None (Two point calibration only).
5. Under
Enter Sensor Type And Optional Data
, choose the type of sensor and add the serial
number and any notes needed.
6. Select the
Two-Point Calibration
tab. You need two known concentrations of oxygen to perform
this calibration, one being the 0% oxygen concentration.