6.  Electrode care and maintenance

Storage

For best results, keep the pH bulb wet. Store the pH bulb in the electrode soaker 
bottle filled with electrode storage solution. Or you can store the electrode in a 
pH 4 buffer with 1/100 part saturated KCl. Other pH buffers are also suitable for
storage. NEVER use deionized water for storage.

After measuring

1. 

Rinse the electrode and reference junction in deionized water.

2. 

Store the electrode as recommended in "Storage" or as recommended 
by the manufacturer.

Electrode cleaning

Because your pH electrode is susceptible to dirt and contamination, clean it every
one to three months depending on extent and condition of use. 

Clean the electrode in a mild detergent solution. Wipe the probe with a soft tissue
paper. Avoid touching the glass membrane with your fingers. Wash thoroughly 
in tap water and then in distilled water. Recalibrate your meter after cleaning 
the electrode.

Special Cleaning Tips

Salt deposit: 

dissolve the deposit by immersing the electrode in tap water for ten to

fifteen minutes. Then thoroughly rinse with distilled water.

Oil/grease film: 

wash electrode pH bulb gently in detergent solution. Rinse elec-

trode tip with distilled water.

Clogged reference junction: 

heat a diluted KCl solution for 60-80°C. Place the sens-

ing part of the electrode into the heated solution for about 10 minutes. Allow the
electrode to cool in some unheated KCl solution.

Protein deposits:

prepare a 1% pepsin solution in 0.1M of HCl. Set the electrode in

the solution for five to ten minutes. Rinse the electrode with distilled water.

7.  Troubleshooting

Problem

Cause

Solution

Err flashes 

Wrong or contaminated 

Use fresh buffer

on LCD

buffer used during 

solution: pH 4.01, 7.00,

calibration

or 10.00

Bad electrode

Change electrode

“Ur” or “Or” on LCD  

pH out of range

with pH  or °C 

Temperature out of range

Cool/Heat sample as needed

annuciator

Bad electrode

Change electrode

Bad temperature

Change temperature

sensor

sensor 

LO annuciator 

Low battery

Replace batteries

on LCD

with fresh set

Power on but 

Batteries not

Insert batteries

no display

in place

Batteries not in 

Re-insert batteries

correct polarity (+ and -) 

with correct polarity

Weak batteries

Replace batteries

Unstable

Electrode not deep

Place electrode 

reading

enough in sample

deeper in sample

Insufficient 

Fill electrode with

reference electrolyte 

reference electrolyte

in electrode

(if electrode is refillable);
replace electrode.

Broken electrode

Replace electrode

External “noises” or

Remove or switch off 

induction caused by

interfering motor

by nearby electric motor

Dirty electrode

Clean electrode

Slow response

Dirty electrode

Clean electrode

6.1

6.2

6.3

Summary of Contents for pH 5+

Page 1: ...DISTRIBUTED BY 00702 41 OPERATING INSTRUCTIONS OAKTON WD 35613 00 01 10 11 pH 5 and pH 6 Acorn Series Meters Printed in the U S A 8 98 pH 6 Acorn series pH mV C Meter ON OFF CAL HOLD ENTER MODE INC...

Page 2: ...ng the Meter on 7 4 Calibration 8 10 4 1 pH Calibration 8 9 4 2 Temperature Calibration 9 4 3 Millivolt mV Calibration 10 5 Measurement 11 5 1 Taking Measurements 11 5 2 HOLD feature 11 6 Electrode Ca...

Page 3: ...ectrode maintenance error messages and troubleshooting This part of the manual also includes the Specifications Accessories Warranty and Return of Items section 2 Display and Keypad Functions The Acor...

Page 4: ...e electrode s connector push together and twist the electrode connector 1 2 turn until it clicks into place To remove simply rotate the connector counter clockwise until it unlocks and slide the conne...

Page 5: ...to indicate CAlibration See figure The display will then show the current noncalibrated reading blinking while in calibration mode See figure 5 Allow the reading to stabilize The meter auto matically...

Page 6: ...le value and then loops back to the minimum allowable value 4 Press ENTER key to confirm calibration The LCD displays CO and the meter then reverts to measurement mode 5 Measurement Taking Measurement...

Page 7: ...heat a diluted KCl solution for 60 80 C Place the sens ing part of the electrode into the heated solution for about 10 minutes Allow the electrode to cool in some unheated KCl solution Protein deposi...

Page 8: ...e and temperature probe WD 35613 11 Additional pH 6 Series Acorn Meter pH mV C with temperature probe electrode not included Accessories WD 35606 80 Protective Rubber Boot encases meter in sturdy rubb...

Page 9: ...to be free from significant deviations in material and workmanship for a period of six months from date of purchase If repair or adjustment is necessary and has not been the result of abuse or misuse...

Page 10: ...19 18...

Reviews: