3. Preparation

Inserting the batteries

The battery compartment is found at the
back of the instrument. To open the 
battery compartment, push the lid up. 

See figure

Note the polarity of battery before insert-
ing the batteries into position. After
replacement, place the cover back into 
its position and press down until it 
locks tight.

A "LO" annunciator in the LCD alerts
you when battery power is running 
low. Replace batteries with a fresh set 
as soon as possible.

Connecting the Electrode and Temperature Probe

To connect the Acorn meter to your electrode, align the post of the meter's connector
with the slots on the electrode's connector, push together and twist the electrode 
connector 1/2 turn until it clicks into place. To remove, simply rotate the connector
counter-clockwise until it unlocks, and slide the connector off the socket.

Insert the phono jack of the temperature sensor into the socket on the meter. 
Unplug the phono jack when not in use.

7

6

3.1

A

A

electrode and temperature probe connections at top of Acorn meter

3.2

Switching the meter on

Press the ON/OFF key to power up your meter. All
the LCD segments display for a few seconds as the
meter goes through a self-diagnostic test. The LCD
then switches into pH measurement mode.

If the LCD then displays "Ur"., the electrode is
faulty, the temperature sensor is faulty, or there is an
open circuit. 

See figure 

See page 13 for more troubleshooting information. 

3.3

A

A

Summary of Contents for pH 5+

Page 1: ...DISTRIBUTED BY 00702 41 OPERATING INSTRUCTIONS OAKTON WD 35613 00 01 10 11 pH 5 and pH 6 Acorn Series Meters Printed in the U S A 8 98 pH 6 Acorn series pH mV C Meter ON OFF CAL HOLD ENTER MODE INC...

Page 2: ...ng the Meter on 7 4 Calibration 8 10 4 1 pH Calibration 8 9 4 2 Temperature Calibration 9 4 3 Millivolt mV Calibration 10 5 Measurement 11 5 1 Taking Measurements 11 5 2 HOLD feature 11 6 Electrode Ca...

Page 3: ...ectrode maintenance error messages and troubleshooting This part of the manual also includes the Specifications Accessories Warranty and Return of Items section 2 Display and Keypad Functions The Acor...

Page 4: ...e electrode s connector push together and twist the electrode connector 1 2 turn until it clicks into place To remove simply rotate the connector counter clockwise until it unlocks and slide the conne...

Page 5: ...to indicate CAlibration See figure The display will then show the current noncalibrated reading blinking while in calibration mode See figure 5 Allow the reading to stabilize The meter auto matically...

Page 6: ...le value and then loops back to the minimum allowable value 4 Press ENTER key to confirm calibration The LCD displays CO and the meter then reverts to measurement mode 5 Measurement Taking Measurement...

Page 7: ...heat a diluted KCl solution for 60 80 C Place the sens ing part of the electrode into the heated solution for about 10 minutes Allow the electrode to cool in some unheated KCl solution Protein deposi...

Page 8: ...e and temperature probe WD 35613 11 Additional pH 6 Series Acorn Meter pH mV C with temperature probe electrode not included Accessories WD 35606 80 Protective Rubber Boot encases meter in sturdy rubb...

Page 9: ...to be free from significant deviations in material and workmanship for a period of six months from date of purchase If repair or adjustment is necessary and has not been the result of abuse or misuse...

Page 10: ...19 18...

Reviews: