background image

9

8

4. Calibration

pH calibration

This instrument is capable of up to three-point calibration to ensure accuracy across
the entire range of the meter. You can perform 1-, 2- or 3-point calibration with 
standard pH buffers 4.01; 7.00; and 10.00.

We recommend you perform at least a 2-point calibration at room temperature using
standard buffers that bracket (one above and one below) the expected sample range.
You can also perform a 1-point calibration, but make sure that the buffer value is
close to the sample value you are measuring.

All new calibrations will over-ride existing data. 

NOTE

Do not reuse buffer solutions after calibration. Contaminants in the solu-

tion can affect the calibration, and the accuracy of the measurements.

Before use: 

remove the electrode soaker bottle. If the electrode has been stored dry,

condition the glass bulb by soaking it in tap water for 30 minutes. This hydrates the
glass bulb if the electrode is too dry, or has not been used for a long period of time.

1.

Turn meter on. Meter will automatically enter pH measurement mode.

2.

Rinse electrode thoroughly with deionized water or a rinse solution. 
DO NOT wipe the electrode;

this causes a build-up of electrostatic 

charge on the glass surface.

3. 

Dip both the electrode  and temperature sensor into pH 7.00 buffer solution. The
glass bulb must be completely immersed into the sample. Stir gently, and wait for
reading to stabilize (approximately 40 seconds). 

4. 

Press CAL key to enter the calibration mode. 
The display will momentarily flash “CA” to
indicate CAlibration.

See figure

The display will then show the current 
noncalibrated reading, blinking while in 
calibration mode.

See figure

5. 

Allow the reading to stabilize. The meter auto-
matically recognizes 7.00, 4.01 or 10.00 buffers.

6. 

Press ENTER key once to confirm calibration.
The LCD displays “CO” to indicate the calibra-
tion point has been confirmed. The meter exits
calibration mode and returns to measurement
mode.

7. 

Repeat with pH buffers 4.01 and/or 10.00 
for best accuracy.

NOTE

: This meter has automatic buffer recogni-

tion which identifies the correct pH buffer values
during calibration. If buffers other than 4.01, 7.00
or 10.00 are used, or the electrode has worn out,
the LCD will flash "Er1".

See figure

Temperature calibration

The temperature sensor included with your meter is factory calibrated. Over time,
the temperature calibration may drift and require recalibration. If you replace the
probe you should calibrate temperature prior to pH or mV calibration.

1. 

Connect your temperature probe to the meter.

2. 

Press the MODE/INC key until "°C" appears in the LCD.

3. 

Compare displayed value to a NIST certified thermometer or other thermometer
known to be accurate. For best accuracy, place probe and thermometer in a 
constant temperature bath.

4. 

Press the CAL key. the LCD shows "CA" and
the reading flashes. 

See figures

5. 

Press MODE/INC key until the display
shows the correct temperature. The MODE
/INC key will scroll to the maximum allow-
able value, and then loop back to the mini-
mum allowable value adjustment (maximum
adjustment is ±5°C from factory default).

6. 

Press ENTER key to confirm calibration.
The LCD displays "CO", and the meter then
reverts to measurement mode.

4.1

A

B

pH

A

B

C

C

4.2

A

 C

B

A

B

Summary of Contents for pH 5+

Page 1: ...DISTRIBUTED BY 00702 41 OPERATING INSTRUCTIONS OAKTON WD 35613 00 01 10 11 pH 5 and pH 6 Acorn Series Meters Printed in the U S A 8 98 pH 6 Acorn series pH mV C Meter ON OFF CAL HOLD ENTER MODE INC...

Page 2: ...ng the Meter on 7 4 Calibration 8 10 4 1 pH Calibration 8 9 4 2 Temperature Calibration 9 4 3 Millivolt mV Calibration 10 5 Measurement 11 5 1 Taking Measurements 11 5 2 HOLD feature 11 6 Electrode Ca...

Page 3: ...ectrode maintenance error messages and troubleshooting This part of the manual also includes the Specifications Accessories Warranty and Return of Items section 2 Display and Keypad Functions The Acor...

Page 4: ...e electrode s connector push together and twist the electrode connector 1 2 turn until it clicks into place To remove simply rotate the connector counter clockwise until it unlocks and slide the conne...

Page 5: ...to indicate CAlibration See figure The display will then show the current noncalibrated reading blinking while in calibration mode See figure 5 Allow the reading to stabilize The meter auto matically...

Page 6: ...le value and then loops back to the minimum allowable value 4 Press ENTER key to confirm calibration The LCD displays CO and the meter then reverts to measurement mode 5 Measurement Taking Measurement...

Page 7: ...heat a diluted KCl solution for 60 80 C Place the sens ing part of the electrode into the heated solution for about 10 minutes Allow the electrode to cool in some unheated KCl solution Protein deposi...

Page 8: ...e and temperature probe WD 35613 11 Additional pH 6 Series Acorn Meter pH mV C with temperature probe electrode not included Accessories WD 35606 80 Protective Rubber Boot encases meter in sturdy rubb...

Page 9: ...to be free from significant deviations in material and workmanship for a period of six months from date of purchase If repair or adjustment is necessary and has not been the result of abuse or misuse...

Page 10: ...19 18...

Reviews: