matic HOT end user manual
6
1.3
Safety
The O
2
matic HOT has a failsafe rechargeable battery that keeps the device running
upon short power breaks. To shut down the O
2
matic HOT press and hold
for 4
seconds. Upon intended shutdown, or unintended shutdown, such as total loss of
power, the device maintains its last settings.
In case of malfunction it reverts to the defined safe-mode, where the user is informed
by turning on the warning icon and an audible alarm.
In safe-mode the flow of oxygen is reduced to zero. The patient shall change to an
alternative oxygen supply and contact the service provider.
As part of the treatment the patient will be assigned a dedicated tablet which will have
the O
2
matic HOT app installed. The app will inform the patient of any relevant
notifications and display information from the HOT device. During treatment it is
advised not to turn off the dedicated tablet.
1.4
Disclaimer
O
2
matic cannot assume responsibility for patient safety, upon:
•
Any sign or evidence of opening the O
2
matic HOT, except by authorized
personnel.
•
Use of any unapproved pulse oximeters.
•
This IFU has been issued in compliance with requirements set out in Council
Directive 93/42 EC concerning Medical Devices.
•
All liability is discontinued if the product is not used as intended and as
described in this IFU.