D 8.4.1.10
11
1. FLOW INDICATOR
Shows the amount of oxygen currently being delivered. If the amount of
oxygen the device is delivering is different from the expected amount, the
device will light up two values. One value is blinking which refers to the
expected amount of oxygen. The other will light up statically, which
indicates how much oxygen is currently being delivered.
Upon start up, two values are lit for 3 seconds, indicating the flow
limitations for current device configuration.
2. OXYGEN SOURCE ERROR
Check oxygen source.
3. NOTIFICATIONS
Check your App for more information.
4. BATTERY LEVEL
Green blinking: Charging
Green: Battery status good
Yellow: Battery status low. Connect to power source when possible
Red: Battery status critical. Connect to power source immediately
Red blinking: Battery fail. Contact service provider
Function of the device is not affected by the device being connected to
power or not.
5. NO STABLE POWER SOURCE
Connect to stable power source when possible.
6. SATURATION MEASUREMENT
Constant lid: Put pulse oximeter on finger. Treatment will start.
Blinking: Signal is lost, reconnect be adjusting your finger.
Off: Session finished. Take pulse oximeter off finger.
7. OXYGEN OUTPUT
Connect your oxygen mask here.
8. ON/OFF BUTTON
The button will light up when the device is turned on.