O
2
matic HOT end user manual
4
1 Warnings, Cautions and Symbols
1.1
Warnings
a)
Use only pulse oximetry sensors that are approved by O
2
matic. Contact your
service provider for more information on approved pulse oximetry for use with
the O
2
matic HOT. Use of any other pulse oximeters increases the risk for
compromising patient safety and will void the warranty.
b)
Do not use in an explosive atmosphere or in the presence of flammable
anesthetics or gases.
c)
Refer to the applicable sensor instructions for use for additional warnings and
cautions.
d)
Regularly check the battery indicator. If lit, see section 5.3 for battery
instructions.
e)
O
2
matic HOT is only to be used with pure oxygen.
f)
Only use accessories such as masks, catheters and hoses approved by you
service provider
.
g)
Do not open the device under any circumstances. Do not use any tools on the
device.
h)
Use of this equipment adjacent to or stacked with other equipment should be
avoided because it could result in improper operation. If such use is necessary,
this equipment and the other equipment should be observed to verify that
they are operating normally.
i)
Use of accessories, transducers and cables other than those specified or
provided by the manufacturer of this device could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
j)
Portable RF (Radio frequency) communications equipment (including
peripherals such as antenna cables and external antennas) should be used no
closer than 30 cm (12 inches) to any part of the O
2
Matic HOT device, including
Warning
Indicates that you must be extremely careful when executing these
instructions. Not complying with these warnings can cause serious injuries
and even death.
Summary of Contents for HOT 100