Section 3
Disassembly of the Track Mount before Installation
Keep the side panels and dress covers in a safe place to prevent damage until
reassembly.
•
Remove the two stop pins P/N 1001076 and the #6
flat head screws P/N 0001155 holding
the pins in place from the track end with the end plate already removed. See figure 27.
Secure all screws and hardware until it is time for reassembly.
Disconnect the Molex connector attached to the trolley first.
•
Slide the trolley assembly P/N 1000187 off the track. See figure 28.
Stop
Pins
Flat
Heads
FIGURE 27: Stop Pins Removal
Trolley
Assembly
FIGURE 28: Trolley Assembly Removal