Turn-off the equipment and plug it off prior to each maintenance, cleaning and sterilization. 

 

7.1  PERIODIC MAINTENANCE 

  Before every use, check the rotor, inserts, adapters and motor shaft against corrosion, 

deformation and cracks. 
 

  Before loading the rotor, be sure that there is no substance in the tube inserts and the chamber. 

 

 

Any substance left in the tube inserts will break the tubes and any substance left in 
the chamber will damage the rotor and the chamber. 

 

7.2  CLEANING  

  Before cleaning the device on / off switch, unplug the appliance off. 

  Do not use alkaline cleaning reagents and metal brushes for cleaning. Chamber, rotor and the 

accessories should be cleaned immediately if any spillage occurs.  

 
Centrifuge  
 

  Clean the centrifuge body and the chamber regularly with a damp cloth, using soap or a mild 

detergent and dry well after cleaning. 

  The chamber should be cleaned with 70% alcohol once a week. 

 

 

 

In case of any spillage or tube breakage, the chamber should be cleaned well. 

 

 

 
Take necessary precautions while cleaning the chamber after tube breakage. There 
could be glass particles in the chamber which may cut the hands. 
 

 

Rotor, tube inserts and adapters 
 

Dismantling and Reinstalling the Rotor, 
 

  Loosen and take out the bolt fixing the rotor. 

  Hold the rotor from two sides and pull out. 

  While re-

installing the rotor, make sure that it is properly placed on the motor’s axis, such that 

the inscriptions on it are facing the top. 

  Tighten the fixing bolt back. 

  Rotor, tube inserts and adapters should be cleaned with a damp cloth using soap or mild 

detergent ideally every day but at least once a week against corrosion.  

  Rotor, tube inserts and adapters can be washed with a damp cloth using soap or mild 

detergent. 

  For the holes in the rotor, tube inserts and adapters, small and soft nylon brush can be used. 

  Rotor, tube inserts and adapters should be cleaned with 70% alcohol once a week. 

  Rotor, tube inserts and adapters should be dried well after cleaning with a dry, absorbent, non-

wool soft tissue or with paper. You may also use a hair drier for that purpose. 

Summary of Contents for NF 048

Page 1: ...0 N VE SANAY MALZEMELER MALAT VE T CARET A NF 048 BENCH TOP CENTRIFUGE USER S MANUAL Z14 K 25 225 Rev No 06 Rev Date 01 2016...

Page 2: ...lity in the case where the goods are not used in accordance with their proper intent 4 The warranty may not be claimed for damages occurred during the shipment for damages resulting from improper hand...

Page 3: ...ove the warning labels do not operate damaged instrument do not operate the instrument with a damaged cable do not move the instrument during operation In case of a problem contact your N ve agent for...

Page 4: ...PORT 7 4 2 UNPACKING 7 4 3 POSITIONING 8 4 4 MAINS SUPPLY 8 SECTION 5 OPERATING PRINCIPLES 9 5 1 CONTROL PANEL 9 SECTION 6 OPERATING PRINCIPLES 10 6 1 LOADING 10 6 2 PROGRAMMING 11 6 3 MANUAL LID OPEN...

Page 5: ...e microprocessor control system speed RPM or RCF and time could be programmed and the pulse function provides the operator to run the programs of short duration It ensures safe processes by means of t...

Page 6: ...tep 10 rpm Timer set range 1 99 minutes and hold position Timer set step 1 minute Motor Induction Motor Supply Values 230 V 50 60 Hz Power Consumption 450 W External Dim s WXDXH mm 275x355x240 Packing...

Page 7: ...anyone and any hazardous materials should not be in the 300 mm safety zone while centrifuge is running Do not move the device while it is running Do not open the lid while rotor is spinning Apply the...

Page 8: ...the rotor and motor shaft SECTION 3 SYMBOLS Symbol in the operating instructions Attention general hazard area This symbol refers to safety relevant warnings and indicates possibly dangerous situation...

Page 9: ...should not be in the 300 mm security area ATTENTION The centrifuges are designed to operate safely under the following conditions Indoor use only Ambient temperature 5 C to 25 C Maximum relative humid...

Page 10: ...also shown on this display 02 Time Display This display shows the time during programming and the run 03 Lid Key and Led This key opens the lid if the led is on It is active before starting the centri...

Page 11: ...ht as the one diametrically opposite for balancing In case the number of tubes which will be centrifuged is less than the capacity of the rotor the tubes must be placed in opposite inserts If the numb...

Page 12: ...kapa anahtar yla bekleme konumuna getirdikten sonra ve yapt n z her program n al mas bittikten sonra yeni bir program al t rmak i in kapa a p kapaman z gerekmektedir 6 3 MANUAL LID OPENING In case of...

Page 13: ...or tube breakage the chamber should be cleaned well Take necessary precautions while cleaning the chamber after tube breakage There could be glass particles in the chamber which may cut the hands Roto...

Page 14: ...entially infectious Before doing any servicing on the instrument it is very important to thoroughly disinfect all possibly contaminated parts Before the instrument is removed from the laboratory for d...

Page 15: ...tor overheat failure Please wait for the motor to cool down and start the centrifuge again Err 6 Motor driver PCB is defective Lid open This failure occurs when lid is opened during the centrifugation...

Page 16: ...15 SECTION 10 ELECTRICAL CIRCUIT DIAGRAM 10 1 NF 048 ELECTRICAL CIRCUIT DIAGRAM...

Page 17: ...16 SECTION 11 WARNING LABEL GROUNDED PLUG...

Reviews: