  In case of any spillage or tube breakage, the rotor, the tube inserts and the adapters should be 

washed well. Special attention should be paid to the glass particles which may be left in the 
tube inserts and adaptors. Be sure that those kinds of particles are removed and cleaned well 
because the remaining particles may cause further tube breakage. 

 

7.3  STERILIZATION 

Rotors, tube inserts and adapters shall be left in contact with alcohol (70% ethanol or izopropanol) 
at least for 10 minutes for sterilization purposes against bacteria and viruses. 
 
Rotors, tube inserts and adapters can be autoclaved at 215 kPa absolute pressure and 121°C for 
20 minutes against microorganisms. 
  

  Do not use formaldehyde - which is poisonous - for sterilization. 

  Do not use phenol - which is corrosive - for sterilization. 

  Do not use gluteraldehide - which is toxic - for sterilization. 

 
 
 
 

 

 

Instruments  or  parts  which  may  come  in  contact  with  biological  samples  (patient 
specimens, controls etc.) should be considered at least potentially infectious. 
Before doing any servicing on the instrument it is very important to thoroughly disinfect all 
possibly  contaminated  parts.  Before  the  instrument  is  removed  from  the  laboratory  for 
disposal or servicing, it must be decontaminated/disinfected. Decontamination/disinfection 
should  be  performed  by  well  trained  personnel,  observing  all  necessary  safety 
precautions. 

 

 

SECTION 8 

DISPOSAL MANAGEMENT CONCEPT 

The currently valid local regulations governing disposal must be observed. It is in the responsibility 
of the user to arrange proper disposal of the individual components. Applicable local regulations for 
disposal  have  to  be  carefully  observed.  The  instruments  and  electronic  accessories  (without 
batteries,  power  packs etc.) must be disposed off according to the regulations for the disposal of 
electronic  components.  Batteries,  power  packs  and  similar  power  source  have  to  be  dismounted 
from electric/electronic parts and disposed off in accordance with applicable local regulations.  

SECTION 9 

TROUBLESHOOTING 

If the centrifuge fails to operate, check the following, 
 

  The on/off switch is on, 

  The mains supply is present, 

  The plug is plugged-in properly, 

  The plug is not defective, 

Summary of Contents for NF 048

Page 1: ...0 N VE SANAY MALZEMELER MALAT VE T CARET A NF 048 BENCH TOP CENTRIFUGE USER S MANUAL Z14 K 25 225 Rev No 06 Rev Date 01 2016...

Page 2: ...lity in the case where the goods are not used in accordance with their proper intent 4 The warranty may not be claimed for damages occurred during the shipment for damages resulting from improper hand...

Page 3: ...ove the warning labels do not operate damaged instrument do not operate the instrument with a damaged cable do not move the instrument during operation In case of a problem contact your N ve agent for...

Page 4: ...PORT 7 4 2 UNPACKING 7 4 3 POSITIONING 8 4 4 MAINS SUPPLY 8 SECTION 5 OPERATING PRINCIPLES 9 5 1 CONTROL PANEL 9 SECTION 6 OPERATING PRINCIPLES 10 6 1 LOADING 10 6 2 PROGRAMMING 11 6 3 MANUAL LID OPEN...

Page 5: ...e microprocessor control system speed RPM or RCF and time could be programmed and the pulse function provides the operator to run the programs of short duration It ensures safe processes by means of t...

Page 6: ...tep 10 rpm Timer set range 1 99 minutes and hold position Timer set step 1 minute Motor Induction Motor Supply Values 230 V 50 60 Hz Power Consumption 450 W External Dim s WXDXH mm 275x355x240 Packing...

Page 7: ...anyone and any hazardous materials should not be in the 300 mm safety zone while centrifuge is running Do not move the device while it is running Do not open the lid while rotor is spinning Apply the...

Page 8: ...the rotor and motor shaft SECTION 3 SYMBOLS Symbol in the operating instructions Attention general hazard area This symbol refers to safety relevant warnings and indicates possibly dangerous situation...

Page 9: ...should not be in the 300 mm security area ATTENTION The centrifuges are designed to operate safely under the following conditions Indoor use only Ambient temperature 5 C to 25 C Maximum relative humid...

Page 10: ...also shown on this display 02 Time Display This display shows the time during programming and the run 03 Lid Key and Led This key opens the lid if the led is on It is active before starting the centri...

Page 11: ...ht as the one diametrically opposite for balancing In case the number of tubes which will be centrifuged is less than the capacity of the rotor the tubes must be placed in opposite inserts If the numb...

Page 12: ...kapa anahtar yla bekleme konumuna getirdikten sonra ve yapt n z her program n al mas bittikten sonra yeni bir program al t rmak i in kapa a p kapaman z gerekmektedir 6 3 MANUAL LID OPENING In case of...

Page 13: ...or tube breakage the chamber should be cleaned well Take necessary precautions while cleaning the chamber after tube breakage There could be glass particles in the chamber which may cut the hands Roto...

Page 14: ...entially infectious Before doing any servicing on the instrument it is very important to thoroughly disinfect all possibly contaminated parts Before the instrument is removed from the laboratory for d...

Page 15: ...tor overheat failure Please wait for the motor to cool down and start the centrifuge again Err 6 Motor driver PCB is defective Lid open This failure occurs when lid is opened during the centrifugation...

Page 16: ...15 SECTION 10 ELECTRICAL CIRCUIT DIAGRAM 10 1 NF 048 ELECTRICAL CIRCUIT DIAGRAM...

Page 17: ...16 SECTION 11 WARNING LABEL GROUNDED PLUG...

Reviews: