6.2 PROGRAMMING
At stand-by position,
Push the Set key. The rotor type starts to flash on the speed display.
Choose the rotor type by using the Value Increase and Decrease keys (nC for the microlitre
rotor, HE for the haematocrit rotor).
The programming type (RPM or RCF) starts to flash on the speed display.
Choose the programming type by using the Value Increase and Decrease keys (RPM or RCF).
Push the Set key again. The speed value on the speed display starts to flash.
Set the speed or RCF by pushing the Value Increase and Decrease keys.
Push the Set key again. The time value on the time display starts to flash.
Set the time by pushing the Value Increase and Decrease keys.
Push the Set key again to save the settings to memory.
Santrifüj işlemi sonrasında, kapak açılmazsa, program tekrar çalıştırılamaz. Bir
sonraki çalışma için kapağı açıp kapatmayı unutmayınız.
Cihazı kapağı kapalı iken aç/kapa anahtarıyla bekleme konumuna getirdikten sonra
ve yaptığınız her programın çalışması bittikten sonra yeni bir program çalıştırmak
için kapağı açıp kapamanız gerekmektedir.
6.3 MANUAL LID OPENING
In case of a power cut or any defects, the instrument would be opened manually to be able to get
the samples.
To open the lid manually :
Power off the instrument
Insert the manual lid opening tool into the hole at the front side of the instrument.
Push the tool while keeping it vertically until the lid is opened.
Before opening the lid manually, be sure that the rotor is completely stopped. Observe the
rotor upon opening while the lid is hold by hand, if the rotor is still rotating, close the lid
and wait approximately 10 minutes before repeating the operation. This operation must
be carried by someone who is informed of the danger and of the precautions which must
be undertaken.
SECTION 7
PERIODIC MAINTENANCE, CLEANING AND STERILIZATION
Her periyodik bakım öncesinde cihazı aç / kapa anahtarından kapatarak fişini
çekiniz.
After every care, check the continuity the network of ground line.