11 

 

6.2  PROGRAMMING 

 
At stand-by position, 
 

  Push the Set key. The rotor type starts to flash on the speed display. 

  Choose  the  rotor  type  by  using  the  Value  Increase  and  Decrease  keys  (nC  for  the  microlitre 

rotor, HE for the haematocrit rotor). 

  The programming type (RPM or RCF) starts to flash on the speed display. 

  Choose the programming type by using the Value Increase and Decrease keys (RPM or RCF). 

  Push the Set key again. The speed value on the speed display starts to flash. 

  Set the speed or RCF by pushing the Value Increase and Decrease keys. 

  Push the Set key again. The time value on the time display starts to flash. 

  Set the time by pushing the Value Increase and Decrease keys. 

  Push the Set key again to save the settings to memory. 

  

 

Santrifüj  işlemi  sonrasında,  kapak  açılmazsa,  program  tekrar  çalıştırılamaz.  Bir 
sonraki çalışma için kapağı açıp kapatmayı unutmayınız. 
 

 

Cihazı kapağı kapalı iken aç/kapa anahtarıyla bekleme konumuna getirdikten sonra 
ve yaptığınız her programın çalışması bittikten sonra yeni bir program çalıştırmak 
için kapağı açıp kapamanız gerekmektedir.

 

 

6.3  MANUAL LID OPENING 

 
In case of a power cut or any defects, the instrument would be opened manually to be able to get 
the samples. 
 
To open the lid manually : 
 

  Power off the instrument 

  Insert the manual lid opening tool into the hole at the front side of the instrument. 

  Push the tool while keeping it vertically until the lid is opened. 

 

 

Before opening the lid manually, be sure that the rotor is completely stopped. Observe the 
rotor upon opening while the lid is hold by hand, if the rotor is still rotating, close the lid 
and  wait  approximately  10  minutes  before  repeating  the  operation.  This  operation  must 
be carried by someone who is informed of the danger and of the precautions which must 
be undertaken. 

 

SECTION  7 

PERIODIC MAINTENANCE, CLEANING AND STERILIZATION 

 

Her  periyodik  bakım  öncesinde  cihazı  aç  /  kapa  anahtarından  kapatarak  fişini 
çekiniz. 

 

 
After every care, check the continuity the network of ground line.

 

Summary of Contents for NF 048

Page 1: ...0 N VE SANAY MALZEMELER MALAT VE T CARET A NF 048 BENCH TOP CENTRIFUGE USER S MANUAL Z14 K 25 225 Rev No 06 Rev Date 01 2016...

Page 2: ...lity in the case where the goods are not used in accordance with their proper intent 4 The warranty may not be claimed for damages occurred during the shipment for damages resulting from improper hand...

Page 3: ...ove the warning labels do not operate damaged instrument do not operate the instrument with a damaged cable do not move the instrument during operation In case of a problem contact your N ve agent for...

Page 4: ...PORT 7 4 2 UNPACKING 7 4 3 POSITIONING 8 4 4 MAINS SUPPLY 8 SECTION 5 OPERATING PRINCIPLES 9 5 1 CONTROL PANEL 9 SECTION 6 OPERATING PRINCIPLES 10 6 1 LOADING 10 6 2 PROGRAMMING 11 6 3 MANUAL LID OPEN...

Page 5: ...e microprocessor control system speed RPM or RCF and time could be programmed and the pulse function provides the operator to run the programs of short duration It ensures safe processes by means of t...

Page 6: ...tep 10 rpm Timer set range 1 99 minutes and hold position Timer set step 1 minute Motor Induction Motor Supply Values 230 V 50 60 Hz Power Consumption 450 W External Dim s WXDXH mm 275x355x240 Packing...

Page 7: ...anyone and any hazardous materials should not be in the 300 mm safety zone while centrifuge is running Do not move the device while it is running Do not open the lid while rotor is spinning Apply the...

Page 8: ...the rotor and motor shaft SECTION 3 SYMBOLS Symbol in the operating instructions Attention general hazard area This symbol refers to safety relevant warnings and indicates possibly dangerous situation...

Page 9: ...should not be in the 300 mm security area ATTENTION The centrifuges are designed to operate safely under the following conditions Indoor use only Ambient temperature 5 C to 25 C Maximum relative humid...

Page 10: ...also shown on this display 02 Time Display This display shows the time during programming and the run 03 Lid Key and Led This key opens the lid if the led is on It is active before starting the centri...

Page 11: ...ht as the one diametrically opposite for balancing In case the number of tubes which will be centrifuged is less than the capacity of the rotor the tubes must be placed in opposite inserts If the numb...

Page 12: ...kapa anahtar yla bekleme konumuna getirdikten sonra ve yapt n z her program n al mas bittikten sonra yeni bir program al t rmak i in kapa a p kapaman z gerekmektedir 6 3 MANUAL LID OPENING In case of...

Page 13: ...or tube breakage the chamber should be cleaned well Take necessary precautions while cleaning the chamber after tube breakage There could be glass particles in the chamber which may cut the hands Roto...

Page 14: ...entially infectious Before doing any servicing on the instrument it is very important to thoroughly disinfect all possibly contaminated parts Before the instrument is removed from the laboratory for d...

Page 15: ...tor overheat failure Please wait for the motor to cool down and start the centrifuge again Err 6 Motor driver PCB is defective Lid open This failure occurs when lid is opened during the centrifugation...

Page 16: ...15 SECTION 10 ELECTRICAL CIRCUIT DIAGRAM 10 1 NF 048 ELECTRICAL CIRCUIT DIAGRAM...

Page 17: ...16 SECTION 11 WARNING LABEL GROUNDED PLUG...

Reviews: