6
2.3 GENERAL PRESENTATION
Figure 1
1
Control panel
4
Gasket
2
Manual LID openıng hole
6
Fixing feeling
3
Hematocrit rotor
7
LID bolt
4
Observatıon hole
8
LID
2.4 PRECAUTIONS AND USAGE LIMITATIONS
Do not use the device for any purpose other than the usage purpose.
Prior to first use, the user’s manual should be read and the device is only to be used by
authorized and trained personnel. Only authorized technical personnel handle the product in
case of any failure.
The working bench should be durable to the device weight and vibration isolated.
Ensure that the rotor is placed correctly prior to usage.
According to the standard IEC 61010-2-020, anyone and any hazardous materials should not
be in the 300 mm safety zone while centrifuge is running.
Do not move the device while it is running.
Do not open the lid while rotor is spinning.
Apply the manual lid opening procedure in the case of power cut or in the case of any error.
Use only the the spare parts, rotors and accessories which are supplied by NUVE.
Load the rotor
according to the explanations in the user’s manual.
Start the device after ensuring the rotor is loaded correctly.
Do not use the centrifuge in areas which are in explosive danger.
Do not centrifuge the explosive, flammable, radioactive, corrosive materials and the materials
which may react with each other.
The centrifuge and the rotor are not microbiologically leak-proof. Use tubes with leak-proof
covers, if hazardous, toxic and pathogenic microorganisms are centrifuged.