background image

Smart Communication Module

7/24

D24        

 

 

59,65

48

3. 

TECHNICAL INFORMATION

3.1 

Technical specification

The SCM-1000 complies with the following technical specification:

Table 1: Technical data of the

 SCM-1000

3.2 

Technical drawing

Illustration 1: Technical drawing (all dimensions in mm)

Parameter

SCM

Unit

Interfaces

USB 2.0

WiFi 802.11 b/g/n

Supply voltage

5.0 ± 10%

V

Current consumption (max.)

165

mA

Samplerate Position (max.)

100

Hz

Weight

50

g

IP-rating (EN 60 529)

IP50

Dimensions 

59.65 x 24

mm

Operating temperature

0 to +55 (+32 to +131)

°C (°F)

Storage temperature

-20 to +70 (-4 to +158)

°C (°F)

Shock [per 11 ms]

≤ 500 (EN 60 068-2-27)

m/s

2

Vibration [within 55-2000 Hz]

≤ 100 (EN 60 068-2-6)

m/s

2

Range (typical)

50

m

Summary of Contents for 1232294-02

Page 1: ...USER MANUAL English Original Instruction SMART COMMUNICATION MODULE FOR LENGTH GAUGE ACANTO...

Page 2: ...rawing 7 3 3 Nameplate 8 4 STATUS MESSAGES IN OPERATING MODE 9 5 INITIAL SETUP OF THE SMART COMMUNICATION MODULE 10 5 1 Factory Settings 10 5 2 Mechanical connections 10 5 3 Compatible software 11 5 4...

Page 3: ...and other applications This standard is maintained and developed by the OPC UA FOUNDATION Essential features Wireless data transmission up to 50 m Transmission of measured values diagnostic data and e...

Page 4: ...all warranty claims NUMERIK JENA GmbH does not assume any liability for any damages or operating errors caused by incorrect mounting and or startup operations Please consider user manuals data sheets...

Page 5: ...difications not expressly approved by the party responsible for compliance could void the user s authority to operate the equipment This device has been tested and found to comply with the limits for...

Page 6: ...ntal Protection and Disposal Environmental damage due to incorrect disposal of the product accessories or peripherals Do not dispose in domestic waste Dispose only by authorized collection points Elec...

Page 7: ...echnical drawing all dimensions in mm Parameter SCM Unit Interfaces USB 2 0 WiFi 802 11 b g n Supply voltage 5 0 10 V Current consumption max 165 mA Samplerate Position max 100 Hz Weight 50 g IP ratin...

Page 8: ...Smart Communication Module 8 24 3 3 Nameplate Illustration 2 Figure of the SCM 1000 nameplate Product Name FCC Declaration Identity Number Serial Number Data Matrix Code...

Page 9: ...onnector to connect the length gauge USB connector type Micro B USB power supply Status LED 4 STATUS MESSAGES IN OPERATING MODE The current status of the SCM 1000 is indicated by the integrated LED ch...

Page 10: ...these factory defaults will be loaded and a hotspot will be generated automatically by the SCM in case no length gauge ACANTO is connected Table 3 Overview of the factory settings 5 2 Mechanical conne...

Page 11: ...s the Smart Communication App SCA as free of charge software for Windows and Android The SCA includes the following functions SCM MANAGER management assistant for the integrated SCMs via WiFi or USB A...

Page 12: ...lustration 8 Figure SCA Menu 5 3 1 SCM in WiFi mode OPC UA compatible If the SCM 1000 is used in WiFi mode a device with an installed OPC UA client is required For the following operating systems NUME...

Page 13: ...9 SCM with WLAN LED lights up red Connect the SCM 1000 to your setup device directly via the USB interface or via WiFi For the WiFi connection an access point is automatically established by the SCM 1...

Page 14: ...t 0 WLAN Client 1 WLAN Access Point 2 Open the Smart Communication App SCA from NUMERIK JENA GmbH to set up the network settings If you are using an alternative software for setup please refer to Tabl...

Page 15: ...nge the customer specific settings the length gauge must be disconnected from the SCM and restarted by interrupting the power supply Illustration 14 Schematic diagram SCM LED lights red WLAN access po...

Page 16: ...ered network settings see Chapter 4 will be loaded automatically SCMs that are connected via a WLAN connection can be used by all output devices PC tablet in the network SCMs that are integrated via t...

Page 17: ...ork as well as the locally connected USB devices In this menu measured values and information e g temperature humidity of the SCM ACANTO length gauge are provided In addition changes to the SCMs can b...

Page 18: ...up SCM as an access point during operation Restart necessary by disconnecting the power supply Reset device error Active SCM settings Error code Acanto internal Figure 19 Schematic illustration of the...

Page 19: ...Path SCM 1000 Hardware Pressure Parameter Condition Min Typ Max Unit Temperature Operating Range operational 40 25 85 C Absolute temperature accuracy 0 65 1 0 C Resolution API output resolution 0 01...

Page 20: ...ring Unit 0 mm deg 1 counts Set power save mode SCM 1000 Axes Axis 1 Encoder 1 isActive 0 switch off 1 switch on Get error status SCM 1000 Axes Axis 1 Encoder 1 Error 0 no error 1 error Read device na...

Page 21: ...TROUBLESHOOTING Connection type Behaviour Potential cause Proposed solution Behavior occurs if NO gauge is connected to the SCM SCM LED is not lit Power supply insufficient Check power supply SCM LED...

Page 22: ...evice to the power supply and adjust the network settings see section 5 4 Check in the SCM settings whether the network setting is set to Client or Access Point SCM lights up blue no measurement data...

Page 23: ...Smart Communication Module 23 24 Figure 21 SCM 1000...

Page 24: ...NUMERIK JENA GmbH Im Semmicht 4 07751 Jena Germany Phone 49 3641 4728 0 E Mail info numerikjena de www numerikjena de Version 04 2020...

Reviews: