![Numerik Jena 1232294-02 User Manual Download Page 21](http://html1.mh-extra.com/html/numerik-jena/1232294-02/1232294-02_user-manual_1717676021.webp)
Smart Communication Module
21/24
Description
Scope of delivery
Order number
Smart Communication
Module
▪
SCM - Smart Communication Module
▪
USB-Cable (USB Typ A/USB Typ B micro)
1232294-02
Replacement USB cable
▪
1.8m
1279537-01
SCA (for Android)
SCA is available for free at the Google Playstore
SCA (for Windows)
SCA is available for free at www.numerikjena.de
7.
APPLICATION PROGRAMMING INTERFACE
For the creation of user-specific software, NUMERIK JENA GmbH will provide information on
the application programming interface on request. The API is also available in the download
area at www.numerikjena.de.
8.
ORDER INFORMATION
Table 11: Overview with description, scope of delivery and order number
9.
SCM & SCA – TROUBLESHOOTING
Connection
type
Behaviour
Potential cause
Proposed solution
Behavior occurs if
NO gauge is
connected to the
SCM.
SCM LED is not lit
Power supply insufficient
Check power supply
SCM LED lights up
red
No error, behaviour is
intended in commissioning
mode
Continue commissioning, the SCM
will set up an access point
(see chapter 4)
SCM LED flashes
red
No error, behaviour is
intended in commissioning
mode
Continue commissioning (see
section 5.4)
SCM LED lights red
- Access Point is not
established
SCM is connected to the
PC
SCM connects via the USB port
> Connection via separate power
supply If the SCM is connected via
USB, no access point is established
(see chapter 5.3.2)
SCM LED is not lit
If the LED does not light
up in any of the possible
connection options, this
may be a defect.
Please contact our support
Table 12: Overview of possible error sources while the SCM-1000 is connected.