Guidance and manufacture’s declaration – electromagnetic immunity
Immunity test
Power frequency
magnetic field IEC
61000-4-8
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
3A/m
3A/m
Floors should be wood, concrete or
ceramic tile. If floor are covered with
synthetic material, the relative humidity
should be at least 30%.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
(50Hz/60Hz)
Electrostatic
IEC 61000-4-2
discharge (ESD)
IEC 60601 test Compliance
level
level
Electromagnetic environment - guidance
The Digital Blood Pressure Monitor is intended for use in the electromagnetic environment
specified below. The customer or the user of Digital Blood Pressure Monitor should assure
that it is used in such an environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
Guidance and manufacture’s declaration – electromagnetic immunity
Immunity test
80 MHz to 2.5
GHz
3A/m
3A/m
80 MHz to 800 MHz
800 MHz to 2.5 GHz
Portable and mobile RF communications
equipment should be used no closer to any
part of the Digital Blood Pressure Monitor,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
Radiated RF
IEC 61000-4-3
IEC 60601 test Compliance
level
level
Electromagnetic environment - guidance
Recommended separation distance
The Digital Blood Pressure Monitor is intended for use in the electromagnetic environment
specified below. The customer or the user of Digital Blood Pressure Monitor should assure
that it is used in such an environment.
35
Immunity test
80 MHz to 2.5
GHz
3A/m
3A/m
survey,a should be less than the
compliance level in each frequency range.b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the Digital Blood Pressure
Monitor is used exceeds the applicable RF compliance level above, the Digital Blood
Pressure Monitor should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the
Digital Blood Pressure Monitor.
Radiated RF
IEC 61000-4-3
IEC 60601 test Compliance
level
level
Electromagnetic environment - guidance
Warranty Information
36
Recommended separation distances between
portable and mobile RF communications equipment and the Digital Blood Pressure Monitor.
output power
of transmitter
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
Separation distance
(m)
according to frequency of transmitter
Pressure Monitor is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the Digital Blood Pressure
Monitor can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the Digital
Blood Pressure Monitor as recommended below, according to the maximum output power
of the communications equipment.
Warranty Information
80 MHz to 800 MHz
0.01
0.1
1
10
100
0.117
0.233
0.369
0.738
1.167
2.333
3.689
7.379
11.667
23.333
80 MHz to 800 MHz