NOVUS AUTOMATION
55/57
* CAUTION with the batteries operation temperature. Extremely high or low temperatures can cause rupture and leakage, causing
damage to the device.
13.1
SENSORS RANGE AND ACCURACY
Sensor
Sensor Minimum
Value
Sensor Maximum
Value
Sensor Resolution
Accuracy (%)
Thermocouples
J
-100.0 °C
-148.0 °F
760.0 °C
1400.0 °F
0.1 °C
0.2 °F
0.15 % (F.R.) ± 0.5 °C
K
-150.0 °C
-238.0 °F
1370.0 °C
2498.0 °F
0.1 °C
0.2 °F
0.15 % (F.R.) ± 0.5 °C
T
-160.0 °C
-256.0 °F
400.0 °C
752.0 °F
0.1 °C
0.2 °F
0.15 % (F.R.) ± 0.5 °C
N
-270.0 °C
-454.0 °F
1300.0 °C
2372.0 °F
0.1 °C
0.2 °F
0.15 % (F.R.) ± 0.5 °C
E
-90.0 °C
-130 °F
720.0 °C
1328.0 °F
0.1 °C
0.2 °F
0.15 % (F.R.) ± 0.5 °C
R
-50.0 °C
-58.0 °F
1760.0 °C
3200.0 °F
0.3 °C
0.5 °F
0.15 % (F.R.) ± 0.5 °C
S
-50.0 °C
-58.0 °F
1760.0 °C
3200.0 °F
0.4 °C
0.7 °F
0.15 % (F.R.) ± 0.5 °C
B
500.0 °C
932.0 °F
1800.0 °C
3272.0 °F
0.4 °C
0.7 °F
0.15 % (F.R.) ± 0.5 °C
Pt100
Pt100
-200.0 °C
-328.0 °F
650.0 °C
1202.0 °F
0.1 °C
0.2 °F
0.15 % (F.R.)
Linear
0 to 50 mV
0.000
50.000
0.003 mV
0.15 % (F.R.)
0 to 5 V
0.000
5.000
0.6 mV
0.15 % (F.R.)
0 to 10 V
0.000
10.000
0.6 mV
0.15 % (F.R.)
0 to 20 mA
0.000
20.000
0.001 mA
0.15 % (F.R.)
4 to 20 mA
4.000
20.000
0.001 mA
0.15 % (F.R.)
Digital Input
Count Mode
0
65535
0.01 % (F.R.)
Internal Sensors
Temperature
(NTC)
-40 °C
-40 °F
125.0 °C
257.0 °F
0.1 °C
0.1 °F
0.15 % (F. E.) ± 0.5 °C
Battery Voltage
3.6
6.5
0.01 V
2 % (F.R.)
External Power
Supply Voltage
10.00
30.00
0.01 V
2 % (F.R.)
*
F.R. = Full Range =
Span
Table 08
– Sensors range and accuracy
Accuracy
: The sensor accuracy reading is measured in relation to the Full Range and is proportional to the maximum measuring range of each
sensor. For example, for a Pt100 sensor, which the
LogBox-BLE
can read in the range of -200 to 650 °C, with a precision of 0.15 %, the accuracy
in °C will be (650 °C - (-200 °C)) * 0.15 % = 1.28 °C.