1.3. Calibration
The following turbidity calibrations can be used:
Ac
= automatic calibration, applicable to Std-program and user programs U1 –
U8.
Max. sensitivity is a setting at 90 % of the max. clearness of the solution.
Min. sensitivity is a setting at 10 % of the max. clearness of the solution.
Pc
= program specific calibration.
As Ac, but max. and min. of the turbidity range must be defined by the
user. For this the calibration function must be used for each individual
program.
Ec
= external calibration:
This is a global calibration for all test programs. The turbidity range can be
set by using the calibration function.
Remember:
The factory setting for the calibration mode is Ac, so that there is no need to use a
standard turbidity solution for calibration.