1. Safety Precautions
1.1. Safety instructions
Read operation manual before starting to work with the instrument. Make sure,
that every operator fully understands every function of the instrument and each of
its components.
Working with solvents and high temperatures does not need to be dangerous when
all safety recommendations are understood and obeyed by the operator
1.2. Instrument Safety
·
When mineral oil is used as a solvent, make sure its boiling point is over the
max. level of the temperature profile of
230°C
. If not,
decrease this level to
the boiling point
of the solvent less 20°C or more.
·
Put the magnetic stirring bar
gently into the glass test tube
, by tilting the
tube.
·
In case a glass test tube would break,
the instrument must be switched off
immediately
. All the solvent in the metal jacket holding the test tube must be
removed, before switching on again.
·
Never use a glass test tube twice.
·
Wear
heat protection gloves
when removing a hot test tube or cleaning a hot
temperature sensor tip.
·
Do not remove any part
from the instrument during operation.
·
If you have
doubts about safety
when a situation occurs that is not
mentioned in the operation manual, contact the manufacturer or its
representative.