10. How to create a test parameter set
The CHEMOTRONIC is providing 8 user defined test programs (U1 - 8)and a
standard test program (STD) for quality assurance applications (Prod).
All programs are identical, when the instrument is leaving the factory with one
exception: The standard program
Std
uses the auto calibration function. The
standard test parameter set corresponds to the recommended settings for accurate
cloud point measurements of resin of good or medium solubility in mineral oils, for
instance PKWF 6/9 or PKWF 6/9 new (mfr: Haltermann).
The standard test parameter set consists of the following segments:
Segment
Duration
Temp.
Speed
Heat up segment
fixed (approx. 4-5
min)
room temp. up to
T1 (230 °C standard)
N1 (1100 rpm
standard)
Hold segment
variable,
normal setting 2.0
min
T1 (230 °C standard) N1 (1100 rpm
standard)
Cool segment 1
fixed (approx. 0.5
min)
T2 (210 °C standard) N2 (800 rpm
standard)
Cool segment 2
fixed (approx. 4-5
min)
T3 (100 °C standard) N2 (800 rpm
standard)
Cool segment 3
fixed (approx.4-5
min)
T3 to room temp.
N3 (1200 rpm
standard)