Novis CairMax User Manual Download Page 9

9

CairMax Duo 

User Guide

 

System Preparation

Alternating Use

When clinically required, connect the control unit (if equipped) to 
the mattress and operate the CairMax Duo in alternating (dynamic) 
mode. 

Once connected, allow up to 20 minutes for the air cells to inflate to 
operational pressure. The patient may remain on the mattress when 
transitioning from Static to Alternating operation. 

1

  Hang the control unit over the foot end of the bed, using the 

inbuilt spring loaded hanging hooks. Ensure it is secure before 
use; failure to do so could result in equipment damage.

2

 Locate the umbilical air hose at the foot end of the mattress 

(it may be stored in the cord retention loops alongside 
the mattress). Connect the umbilical connectors to the 
corresponding sockets on the side of the control unit. Listen for a 
click as confirmation the connector is locked in place.

  Straighten any twists in the umbilical air hose to ensure 

uninterrupted air flow between the control unit and mattress. 
Also ensure  
the air hoses are not trapped between the mattress and bed.  
Failure to do so could result in an under inflated mattress.

3

 Feed power cord through the cord retention loops along the 

side of the mattress base. Insert power cord plug into the side 
of the control unit, then connect to an appropriate electrical 
outlet and switch on mains power.

  Ensure the power cord is not under strain; is free from 

obstruction; and is stored safely so as not to be a trip hazard.

  Periodically inspect the power cord connector for damage  

– the plugs are made of transparent plastic for easy visibility. 
 

4

On the control unit, press and hold the Power button for a 
minimum of three seconds.  

The Power indicator will illuminate green to indicate the system 
is operational. While reaching initial operating pressure, all 
five pressure setting indicators will flash green.  Allow up to 20 
minutes for complete inflation.

 

 Once operating pressure is reached, both Alternate and Auto 
indicators will illuminate to indicate both alternating mode and 
automatic pressure setting are functioning.

3 SEC

Summary of Contents for CairMax

Page 1: ...U S E R G U I D E CairMaxDuo...

Page 2: ...inst foreign object and vertically falling water drops Before operating this medical equipment it is important to read this user guide and understand the operating instructions and safety precautions...

Page 3: ...tem Overview 4 Intended Use 5 Safety Precautions 6 7 System Preparation 8 9 Operation Control Unit 10 11 Operation Mode 12 Operation Transport Mode 12 Operation Storage 12 Troubleshooting 13 Care and...

Page 4: ...e without use of the control unit however when clinically required has the facility to step up to provide dynamic support with the control unit connected The mattress consists of a foam head cell comb...

Page 5: ...sing Shipping Storage Environment Temperature 5 C to 60 C 41 F 140 F Humidity 30 to 75 non condensing Connecting System to Other Devices There no are other devices necessary for normal operation The C...

Page 6: ...oosely so as not to affect the pressure redistrubution As part of a sensible pressure injury prevention strategy avoid wearing clothing that may cause areas of localised damage due to creases seams ob...

Page 7: ...flammable anaesthetic mixture with air oxygen or nitrous oxide Do not use in the presence of smoking materials or open flame air flowing through the mattress will support combustion Do not open the c...

Page 8: ...ver facing upwards and umbilical air hoses towards the left hand base of the bed The mattress can be covered with a light sheet as required Always secure sheets loosely enough to ensure they do not in...

Page 9: ...rupted air flow between the control unit and mattress Also ensure the air hoses are not trapped between the mattress and bed Failure to do so could result in an under inflated mattress 3Feed power cor...

Page 10: ...10 User Guide CairMax Duo Operation Control Unit MAX INFLATE PRESSURE SETTING AUTO SOFT FIRM ALTERNATE STATIC MODE POWER 3 SEC 3 SEC ALARM MUTE A POWER FAIL PRESSURE LOSS PANEL LOCKOUT H C B A D F G E...

Page 11: ...ernative cell sets inflate and deflate in a 12 minute cycle D Power Press and hold for a minimum of 3 seconds to turn system power on and off 3 SEC Green light power on 3 SEC Amber light standby power...

Page 12: ...sure the control unit switched off and disconnected from mains power Disconnect the umbilical air hoses from the control unit by unplugging the quick release connectors wait for all air cells to defla...

Page 13: ...t is making unusual noise Ensure control unit is resting against a solid surface Mattress surface appears to be uneven Ensure all air cells and static head area foam are evenly placed on foam core str...

Page 14: ...in damage to the equipment and may result in damage to the polyurethane coating or negate the biocompatibility properties of the fabric Cleaning and Infection Control It is recommended that the CairMa...

Page 15: ...merse the unit in water or other fluid Disinfection The mattress top cover and control unit may be decontaminated by using a solution of sodium hypochlorite or similar up to 10 000 ppm available chlor...

Page 16: ...s Electromagnetic propagation is affected by absorption and reflection from structures objects and people Guidance and manufacturer s declaration electromagnetic emissions The CairMax Duo is intended...

Page 17: ...gnetic fields should be at levels characteristic of a typical location in a typical home healthcare and professional healthcare environment Conducted RF IEC 61000 4 6 3 Vrms 150 kHz to 80 MHz 3 V m Po...

Page 18: ...Class II Type BF INGRESS PROTECTION RATING IP21 DIMENSIONS HEIGHT 195 mm WIDTH 265 mm DEPTH 120 mm WEIGHT 2 kg MATTRESS DIMENSIONS LENGTH 2000 mm 2000 mm WIDTH 880 mm 1050 mm HEIGHT 160 mm 160 mm WEI...

Page 19: ...option Goods repaired may be replaced by refurbished good of the same type rather than being repaired Refurbished parts may be used to repair goods Novis Healthcare will not be held responsible for an...

Page 20: ...Pressure care and patient handling specialists novis com au 1300 738 885 CMXDUG 0419 APR 2019...

Reviews: