 CairMax Duo

Operation

Mode

ALTERNATE

In Alternating Mode, alternate mattress air cells 
inflate and deflate following a fixed cycle time of 12 
minutes.

STATIC

In Static Mode, all air cells cells remain fully inflated 
to maximum selected pressure setting. This mode 
should be used to create a firm base for stable 
patient handling and transport or other special 
circumstances.

  The system will automatically start in Static 

Mode, and revert to Alternating Mode once 
optimal pressure is reached.

  The system will operate in Static Mode for a 

maximum of  
20 minutes, after which it will automatically 
revert to  
Alternating Mode for patient safety.

Transport Mode

The mattress should be transported in Static Use, 
without control unit.

Before moving the mattress, ensure the control unit 
switched off and disconnected from mains power. 
Disconnect the umbilical air hoses from the control 
unit by unplugging the quick release connectors, wait 
for all air cells to deflate to remove excess pressure. 
Run the umbilical air hoses through the left-side 
cord retention loops and clip the male and female 
connectors around a loop to secure in place.

  The air cells will not alternate when 

disconnected from the control unit. Repeat 
Alternating Use set up instructions (page 9) 
to active dynamic function after mattress is 
moved.

Storage

  Disconnect the mattress from control unit when 

not in use  
and store separately. 

The mattress should be stored laying flat in a cool 
and dry area after cleaning. Avoid laying the mattress 
upside down, or on the side or end. Avoid stacking 
mattresses or placing heavy objects on top of the the 
mattress when not in use.

It is recommended the control unit and power cord be 
returned to its carton when not in use.

Summary of Contents for CairMax

Page 1: ...U S E R G U I D E CairMaxDuo...

Page 2: ...inst foreign object and vertically falling water drops Before operating this medical equipment it is important to read this user guide and understand the operating instructions and safety precautions...

Page 3: ...tem Overview 4 Intended Use 5 Safety Precautions 6 7 System Preparation 8 9 Operation Control Unit 10 11 Operation Mode 12 Operation Transport Mode 12 Operation Storage 12 Troubleshooting 13 Care and...

Page 4: ...e without use of the control unit however when clinically required has the facility to step up to provide dynamic support with the control unit connected The mattress consists of a foam head cell comb...

Page 5: ...sing Shipping Storage Environment Temperature 5 C to 60 C 41 F 140 F Humidity 30 to 75 non condensing Connecting System to Other Devices There no are other devices necessary for normal operation The C...

Page 6: ...oosely so as not to affect the pressure redistrubution As part of a sensible pressure injury prevention strategy avoid wearing clothing that may cause areas of localised damage due to creases seams ob...

Page 7: ...flammable anaesthetic mixture with air oxygen or nitrous oxide Do not use in the presence of smoking materials or open flame air flowing through the mattress will support combustion Do not open the c...

Page 8: ...ver facing upwards and umbilical air hoses towards the left hand base of the bed The mattress can be covered with a light sheet as required Always secure sheets loosely enough to ensure they do not in...

Page 9: ...rupted air flow between the control unit and mattress Also ensure the air hoses are not trapped between the mattress and bed Failure to do so could result in an under inflated mattress 3Feed power cor...

Page 10: ...10 User Guide CairMax Duo Operation Control Unit MAX INFLATE PRESSURE SETTING AUTO SOFT FIRM ALTERNATE STATIC MODE POWER 3 SEC 3 SEC ALARM MUTE A POWER FAIL PRESSURE LOSS PANEL LOCKOUT H C B A D F G E...

Page 11: ...ernative cell sets inflate and deflate in a 12 minute cycle D Power Press and hold for a minimum of 3 seconds to turn system power on and off 3 SEC Green light power on 3 SEC Amber light standby power...

Page 12: ...sure the control unit switched off and disconnected from mains power Disconnect the umbilical air hoses from the control unit by unplugging the quick release connectors wait for all air cells to defla...

Page 13: ...t is making unusual noise Ensure control unit is resting against a solid surface Mattress surface appears to be uneven Ensure all air cells and static head area foam are evenly placed on foam core str...

Page 14: ...in damage to the equipment and may result in damage to the polyurethane coating or negate the biocompatibility properties of the fabric Cleaning and Infection Control It is recommended that the CairMa...

Page 15: ...merse the unit in water or other fluid Disinfection The mattress top cover and control unit may be decontaminated by using a solution of sodium hypochlorite or similar up to 10 000 ppm available chlor...

Page 16: ...s Electromagnetic propagation is affected by absorption and reflection from structures objects and people Guidance and manufacturer s declaration electromagnetic emissions The CairMax Duo is intended...

Page 17: ...gnetic fields should be at levels characteristic of a typical location in a typical home healthcare and professional healthcare environment Conducted RF IEC 61000 4 6 3 Vrms 150 kHz to 80 MHz 3 V m Po...

Page 18: ...Class II Type BF INGRESS PROTECTION RATING IP21 DIMENSIONS HEIGHT 195 mm WIDTH 265 mm DEPTH 120 mm WEIGHT 2 kg MATTRESS DIMENSIONS LENGTH 2000 mm 2000 mm WIDTH 880 mm 1050 mm HEIGHT 160 mm 160 mm WEI...

Page 19: ...option Goods repaired may be replaced by refurbished good of the same type rather than being repaired Refurbished parts may be used to repair goods Novis Healthcare will not be held responsible for an...

Page 20: ...Pressure care and patient handling specialists novis com au 1300 738 885 CMXDUG 0419 APR 2019...

Reviews: