Novis CairMax User Manual Download Page 13

13

CairMax Duo 

User Guide

 

Troubleshooting

SYMPTOM

SOLUTION

Control unit does not operate

Check control unit is connected to the mains power supply.
Check for loose power cord connection and ensure main power is 

switched on.
Check the fuse in rear panel of control unit. Replace if necessary.
Check condition of power cord and plug. Check if mains socket is faulty.

Low pressure indicator 

illuminated; 
mattress is not inflating with 

control unit connected and 

switched on

Ensure the main power is turned on and power cord is connected to mains 

and control unit.
Check control unit/mattress air connections are fitted securely, and 

reconnect umbilical cord if loose.
Ensure control unit is turned on.
Check air intake from filter is not blocked by linen/dust. Replace with new 

filter if needed.
Ensure the air cells are free of damage or leaks

For faster mattress reinflation once the air leak has been closed, press MAX INFLATE and wait until the low 

pressure indicator extinguishes.  

Press the ALTERNATE button to resume alternation

Control unit is making unusual 

noise

Ensure control unit is resting against a solid surface

Mattress surface appears to be 

uneven

Ensure all air cells and static head area foam are evenly placed on foam 

core, straighten and re-lay if uneven, crooked or protruding. 

   If the problem persists, move patient to an alternate product and contact Novis Healthcare.

Waste Disposal

This product has been supplied from an 
environmentally aware manufacturer that complies 
with the European Community’s Waste Electrical and 
Electronic Equipment Directive (WEEE).

This product may contain 
substances that could be 
harmful to the environment if 
disposed of in places (landfills) 
that are not appropriate 
according the legislation.  Please 
be environmentally responsible 
and contact your local authority 
on available options to recycle  
this product at its end of life.

Service Life

The expected service life of a control unit and a 
mattress is highly dependent on frequency of use, 
servicing, care and maintenance.

To maintain the condition of the foam air 
combination mattress system, service the system 
regularly according to the schedule recommended 
by Novis.

  Do NOT use unapproved accessories or attempt 

to modify,  
disassemble or otherwise misuse the CairMax 
Duo system.

Summary of Contents for CairMax

Page 1: ...U S E R G U I D E CairMaxDuo...

Page 2: ...inst foreign object and vertically falling water drops Before operating this medical equipment it is important to read this user guide and understand the operating instructions and safety precautions...

Page 3: ...tem Overview 4 Intended Use 5 Safety Precautions 6 7 System Preparation 8 9 Operation Control Unit 10 11 Operation Mode 12 Operation Transport Mode 12 Operation Storage 12 Troubleshooting 13 Care and...

Page 4: ...e without use of the control unit however when clinically required has the facility to step up to provide dynamic support with the control unit connected The mattress consists of a foam head cell comb...

Page 5: ...sing Shipping Storage Environment Temperature 5 C to 60 C 41 F 140 F Humidity 30 to 75 non condensing Connecting System to Other Devices There no are other devices necessary for normal operation The C...

Page 6: ...oosely so as not to affect the pressure redistrubution As part of a sensible pressure injury prevention strategy avoid wearing clothing that may cause areas of localised damage due to creases seams ob...

Page 7: ...flammable anaesthetic mixture with air oxygen or nitrous oxide Do not use in the presence of smoking materials or open flame air flowing through the mattress will support combustion Do not open the c...

Page 8: ...ver facing upwards and umbilical air hoses towards the left hand base of the bed The mattress can be covered with a light sheet as required Always secure sheets loosely enough to ensure they do not in...

Page 9: ...rupted air flow between the control unit and mattress Also ensure the air hoses are not trapped between the mattress and bed Failure to do so could result in an under inflated mattress 3Feed power cor...

Page 10: ...10 User Guide CairMax Duo Operation Control Unit MAX INFLATE PRESSURE SETTING AUTO SOFT FIRM ALTERNATE STATIC MODE POWER 3 SEC 3 SEC ALARM MUTE A POWER FAIL PRESSURE LOSS PANEL LOCKOUT H C B A D F G E...

Page 11: ...ernative cell sets inflate and deflate in a 12 minute cycle D Power Press and hold for a minimum of 3 seconds to turn system power on and off 3 SEC Green light power on 3 SEC Amber light standby power...

Page 12: ...sure the control unit switched off and disconnected from mains power Disconnect the umbilical air hoses from the control unit by unplugging the quick release connectors wait for all air cells to defla...

Page 13: ...t is making unusual noise Ensure control unit is resting against a solid surface Mattress surface appears to be uneven Ensure all air cells and static head area foam are evenly placed on foam core str...

Page 14: ...in damage to the equipment and may result in damage to the polyurethane coating or negate the biocompatibility properties of the fabric Cleaning and Infection Control It is recommended that the CairMa...

Page 15: ...merse the unit in water or other fluid Disinfection The mattress top cover and control unit may be decontaminated by using a solution of sodium hypochlorite or similar up to 10 000 ppm available chlor...

Page 16: ...s Electromagnetic propagation is affected by absorption and reflection from structures objects and people Guidance and manufacturer s declaration electromagnetic emissions The CairMax Duo is intended...

Page 17: ...gnetic fields should be at levels characteristic of a typical location in a typical home healthcare and professional healthcare environment Conducted RF IEC 61000 4 6 3 Vrms 150 kHz to 80 MHz 3 V m Po...

Page 18: ...Class II Type BF INGRESS PROTECTION RATING IP21 DIMENSIONS HEIGHT 195 mm WIDTH 265 mm DEPTH 120 mm WEIGHT 2 kg MATTRESS DIMENSIONS LENGTH 2000 mm 2000 mm WIDTH 880 mm 1050 mm HEIGHT 160 mm 160 mm WEI...

Page 19: ...option Goods repaired may be replaced by refurbished good of the same type rather than being repaired Refurbished parts may be used to repair goods Novis Healthcare will not be held responsible for an...

Page 20: ...Pressure care and patient handling specialists novis com au 1300 738 885 CMXDUG 0419 APR 2019...

Reviews: