Novis CairMax User Manual Download Page 11

11

CairMax Duo 

User Guide

 

Operation

 - Control Unit

A

   Max Inflate

Rapidly inflates mattress to maximum pressure in 
Static mode. System will automatically return to 
Alternating mode after 20 minutes.

 

B

   Pressure Setting 

Allows manual adjustments to the automatic 
pressure setting (within reasonable limits for the 
patient’s detected weight). Press the button to cycle 
from Soft to Firm pressure. Press the button again to 
cycle back to Auto – green indicator illuminates to 
indicate automatic pressure setting is functioning. 

 

 Mode

Cycle between Static and Alternating modes. Press 
the button to select required mode – indicator glows 
green above the selected mode to indicate the mode 
currently active.

STATIC

In 

Static mode

, all internal 

air cells are inflated to 

the selected pressure 
setting (automatic or 

manual) with no dynamic 

alternation. 

System will automatically 

return to Alternating 

mode after 20 minutes.

ALTERNATE

In 

Alternating 

mode,

 

alternative cell 

sets inflate and 

deflate in a 

12-minute cycle.

 

 Power 

Press and hold for a minimum of 3 seconds to turn 
system power on and off. 

3 SEC

Green light = power on

3 SEC

Amber light = standby power,  

power source connected

  Alarm Mute

Turns audible alarm off temporarily. Press to mute 
the alarm. Alarm will resound in 20 minutes if the 
issues has not been resolved, or immediately if new 
fault detected.

 

F

   Power Failure

During a power failure, amber light flashes and an 
audible alarm sounds to alert carer.  

 

G

   Pressure Loss

Indicates mattress has failed to reach required 
pressure. Indicator flashes amber and after five 
minutes, an audible alarm sounds to alert carers that 
the control unit has failed to reach the set pressure. 
Refer to Troubleshooting for support. 

 

 Lock/Unlock

Lock and unlock the Control Unit panel to prevent 
unwanted interference. 

Press and hold the button for a minimum of three 
seconds – a beep sounds and the light illuminates 
to indicate system is locked. When locked, only 
the Alarm Mute and Lock/Unlock button remain 
operational.

Press again for at least three seconds to unlock (beep 
sounds and light turns off).

Summary of Contents for CairMax

Page 1: ...U S E R G U I D E CairMaxDuo...

Page 2: ...inst foreign object and vertically falling water drops Before operating this medical equipment it is important to read this user guide and understand the operating instructions and safety precautions...

Page 3: ...tem Overview 4 Intended Use 5 Safety Precautions 6 7 System Preparation 8 9 Operation Control Unit 10 11 Operation Mode 12 Operation Transport Mode 12 Operation Storage 12 Troubleshooting 13 Care and...

Page 4: ...e without use of the control unit however when clinically required has the facility to step up to provide dynamic support with the control unit connected The mattress consists of a foam head cell comb...

Page 5: ...sing Shipping Storage Environment Temperature 5 C to 60 C 41 F 140 F Humidity 30 to 75 non condensing Connecting System to Other Devices There no are other devices necessary for normal operation The C...

Page 6: ...oosely so as not to affect the pressure redistrubution As part of a sensible pressure injury prevention strategy avoid wearing clothing that may cause areas of localised damage due to creases seams ob...

Page 7: ...flammable anaesthetic mixture with air oxygen or nitrous oxide Do not use in the presence of smoking materials or open flame air flowing through the mattress will support combustion Do not open the c...

Page 8: ...ver facing upwards and umbilical air hoses towards the left hand base of the bed The mattress can be covered with a light sheet as required Always secure sheets loosely enough to ensure they do not in...

Page 9: ...rupted air flow between the control unit and mattress Also ensure the air hoses are not trapped between the mattress and bed Failure to do so could result in an under inflated mattress 3Feed power cor...

Page 10: ...10 User Guide CairMax Duo Operation Control Unit MAX INFLATE PRESSURE SETTING AUTO SOFT FIRM ALTERNATE STATIC MODE POWER 3 SEC 3 SEC ALARM MUTE A POWER FAIL PRESSURE LOSS PANEL LOCKOUT H C B A D F G E...

Page 11: ...ernative cell sets inflate and deflate in a 12 minute cycle D Power Press and hold for a minimum of 3 seconds to turn system power on and off 3 SEC Green light power on 3 SEC Amber light standby power...

Page 12: ...sure the control unit switched off and disconnected from mains power Disconnect the umbilical air hoses from the control unit by unplugging the quick release connectors wait for all air cells to defla...

Page 13: ...t is making unusual noise Ensure control unit is resting against a solid surface Mattress surface appears to be uneven Ensure all air cells and static head area foam are evenly placed on foam core str...

Page 14: ...in damage to the equipment and may result in damage to the polyurethane coating or negate the biocompatibility properties of the fabric Cleaning and Infection Control It is recommended that the CairMa...

Page 15: ...merse the unit in water or other fluid Disinfection The mattress top cover and control unit may be decontaminated by using a solution of sodium hypochlorite or similar up to 10 000 ppm available chlor...

Page 16: ...s Electromagnetic propagation is affected by absorption and reflection from structures objects and people Guidance and manufacturer s declaration electromagnetic emissions The CairMax Duo is intended...

Page 17: ...gnetic fields should be at levels characteristic of a typical location in a typical home healthcare and professional healthcare environment Conducted RF IEC 61000 4 6 3 Vrms 150 kHz to 80 MHz 3 V m Po...

Page 18: ...Class II Type BF INGRESS PROTECTION RATING IP21 DIMENSIONS HEIGHT 195 mm WIDTH 265 mm DEPTH 120 mm WEIGHT 2 kg MATTRESS DIMENSIONS LENGTH 2000 mm 2000 mm WIDTH 880 mm 1050 mm HEIGHT 160 mm 160 mm WEI...

Page 19: ...option Goods repaired may be replaced by refurbished good of the same type rather than being repaired Refurbished parts may be used to repair goods Novis Healthcare will not be held responsible for an...

Page 20: ...Pressure care and patient handling specialists novis com au 1300 738 885 CMXDUG 0419 APR 2019...

Reviews: