Safety Precautions

Protection Against Hazards

Fluids

Avoid spilling fluids on any part of the control unit. If 
spills do occur:

 

Turn off control unit power and disconnect the 
unit from the mains electricity supply.

 

Immediately clean fluids from the casing by 
wiping with a soft cloth.

  Trapped moisture in foam may lead to an 

infection control hazard. Avoid exposing the 
foam sections to water or liquid.  
Foam is not washable. 

  Ensure there is no moisture in or near the power 

inlet, power switch and power cord before 
reconnecting the power supply.

  Check the operation of controls and other 

components around the spill area.

  Fluid or liquid remaining on the electronic 

controls can cause corrosion that may cause 
the electronic components to fail. Component 
failures may cause the unit to operate erratically, 
possibly producing potential hazards to patient 
and carers.

Explosion hazard

Equipment is not suitable for use in the presence of 
a flammable anaesthetic mixture with air, oxygen or 
nitrous oxide.

 

Do not use in the presence of smoking materials 
or open flame – air flowing through the mattress 
will support combustion.

 

Do not open the control unit – risk of electrical 
shock. Refer servicing to qualified service 
personnel.

Disposal

Dispose of all components (control unit including 
batteries, air filter, air cells, mattress cover and base) 
according to local procedures and regulations or 
contact Novis Healthcare for advice.

Power Cord

The system should never be operated with a worn 
or damaged power cord. Keep the cord away from 
heated surfaces. Should the power cord be found to 
be worn or damaged, contact Novis Healthcare for a 
replacement.

Interference

Although this equipment conforms to the 
intent of directive IEC 60601-1-2

1

 in relation to 

Electromagnetic Compatibility, all electrical 
equipment may produce interference. If interference 
is suspected, move equipment away from sensitive 
devices or contact Novis Healthcare. 

1

 

 IEC 60601-1-2. Medical Electrical Equipment - Part 1: 
General Equipments for Safety, Amendment No. 2. 
Collateral Standard. Electromagnetic Compatibility 
Requirements and Test).

Summary of Contents for CairMax

Page 1: ...U S E R G U I D E CairMaxDuo...

Page 2: ...inst foreign object and vertically falling water drops Before operating this medical equipment it is important to read this user guide and understand the operating instructions and safety precautions...

Page 3: ...tem Overview 4 Intended Use 5 Safety Precautions 6 7 System Preparation 8 9 Operation Control Unit 10 11 Operation Mode 12 Operation Transport Mode 12 Operation Storage 12 Troubleshooting 13 Care and...

Page 4: ...e without use of the control unit however when clinically required has the facility to step up to provide dynamic support with the control unit connected The mattress consists of a foam head cell comb...

Page 5: ...sing Shipping Storage Environment Temperature 5 C to 60 C 41 F 140 F Humidity 30 to 75 non condensing Connecting System to Other Devices There no are other devices necessary for normal operation The C...

Page 6: ...oosely so as not to affect the pressure redistrubution As part of a sensible pressure injury prevention strategy avoid wearing clothing that may cause areas of localised damage due to creases seams ob...

Page 7: ...flammable anaesthetic mixture with air oxygen or nitrous oxide Do not use in the presence of smoking materials or open flame air flowing through the mattress will support combustion Do not open the c...

Page 8: ...ver facing upwards and umbilical air hoses towards the left hand base of the bed The mattress can be covered with a light sheet as required Always secure sheets loosely enough to ensure they do not in...

Page 9: ...rupted air flow between the control unit and mattress Also ensure the air hoses are not trapped between the mattress and bed Failure to do so could result in an under inflated mattress 3Feed power cor...

Page 10: ...10 User Guide CairMax Duo Operation Control Unit MAX INFLATE PRESSURE SETTING AUTO SOFT FIRM ALTERNATE STATIC MODE POWER 3 SEC 3 SEC ALARM MUTE A POWER FAIL PRESSURE LOSS PANEL LOCKOUT H C B A D F G E...

Page 11: ...ernative cell sets inflate and deflate in a 12 minute cycle D Power Press and hold for a minimum of 3 seconds to turn system power on and off 3 SEC Green light power on 3 SEC Amber light standby power...

Page 12: ...sure the control unit switched off and disconnected from mains power Disconnect the umbilical air hoses from the control unit by unplugging the quick release connectors wait for all air cells to defla...

Page 13: ...t is making unusual noise Ensure control unit is resting against a solid surface Mattress surface appears to be uneven Ensure all air cells and static head area foam are evenly placed on foam core str...

Page 14: ...in damage to the equipment and may result in damage to the polyurethane coating or negate the biocompatibility properties of the fabric Cleaning and Infection Control It is recommended that the CairMa...

Page 15: ...merse the unit in water or other fluid Disinfection The mattress top cover and control unit may be decontaminated by using a solution of sodium hypochlorite or similar up to 10 000 ppm available chlor...

Page 16: ...s Electromagnetic propagation is affected by absorption and reflection from structures objects and people Guidance and manufacturer s declaration electromagnetic emissions The CairMax Duo is intended...

Page 17: ...gnetic fields should be at levels characteristic of a typical location in a typical home healthcare and professional healthcare environment Conducted RF IEC 61000 4 6 3 Vrms 150 kHz to 80 MHz 3 V m Po...

Page 18: ...Class II Type BF INGRESS PROTECTION RATING IP21 DIMENSIONS HEIGHT 195 mm WIDTH 265 mm DEPTH 120 mm WEIGHT 2 kg MATTRESS DIMENSIONS LENGTH 2000 mm 2000 mm WIDTH 880 mm 1050 mm HEIGHT 160 mm 160 mm WEI...

Page 19: ...option Goods repaired may be replaced by refurbished good of the same type rather than being repaired Refurbished parts may be used to repair goods Novis Healthcare will not be held responsible for an...

Page 20: ...Pressure care and patient handling specialists novis com au 1300 738 885 CMXDUG 0419 APR 2019...

Reviews: